chiller internal temperature is displayed
• LED
and are on
parameter r9 defines working setpoint
• press
several times:
“PoS” is displayed
- escape this procedure or do not use for 15 seconds
To suspend the cycle:
• press
for 2 seconds:
3.3 Negative temperature timer controlled chilling and storage
To start the cycle:
check that the system is in “stand-by” mode and no other procedure is running
• press
to select “nEg” check that the LED is flashing
press
within 60 seconds: blast chilling time will be displayed (the unit of
measurement is in minutes)
press
or within 15 seconds to modify value setting (setting remains active
until another cycle is selected or when the value defined by parameter
r2 is restored)
• press within 15 seconds: cycle will start running.
If negative temperature blast chilling and storage cycles are not enabled (parameter
rb = 0) tag “nEg” is not displayed.
During blast chilling:
remaining chilling time is displayed
• LED is on
parameter r2 defines chilling time
parameter r8 defines working setpoint
• press
several times:
- “nEg” is displayed
- displays chiller internal temperature
- escape this procedure or do not use for 15 seconds
After chilling time has been reached:
the chiller goes into storage mode
End” is displayed
a buzzer is activated and buzzer time is defined by parameter AA
press a key to turn off the buzzer; press the key again to delete the prompt “End”.
During storage:
chiller internal temperature is displayed
• LED
and are on
parameter rA defines working setpoint
• press
several times: