Manual Supplement 726 Users
4 7/15
Ramp Time
Ramp Time sets the ramp time from 5 sec to 99 sec.
1. Press until RAMP TIME appears on the display.
2. Use and to set the ramp time.
3. Press to save the setting.
On page 25, Table 6, add the following:
Cu10 10 Ω Copper 0.0042 Ω/°C -100 to 250
On page 42, under Auto Ramping the Output, replace the 2
nd
and 3
rd
bullets with the following:
0 % - 100 % - 0 % configurable time smooth ramp. Set ramp time using configuration
menu.
0 % - 100 % - 0 % Stair-step ramp in 25 % steps, pausing at each step. Set ramp
time using configuration menu. Steps are listed in Table 7.
On page 60, under Frequency Measurement, add the following to the bottom of the table:
Sensitivity: 1 V peak to peak minimum
Waveform: Squarewave