The appliance must only be used on a non-slip surface.
Grubby support surface diminish the stability of the appli-
ance.
When working with the shaker, the user must select and
wear his personal protective equipment according to the
mixing hazard category. Defective or inadequate protective
equipment can expose the user to the risk of squirting liquids
and projectile parts. For this reason, only closed test tubes
should be used.
The user may also be at risk if inflammable materials are used
in the shaker, or if the trigger contact pressure is too great
for glass equipment or the transmitted mechanical shaking
power causes the glass to break.
Furthermore, the shaker may only be used to shake mate-
rials or material mixtures that the user knows will not react
dangerously to the extra energy produced by the shaking.
This also applies to extra energy generated by solar radiation
during the shaking procedure.
The appliance is maintenance-free and must not be opened.
The IKA lab dancer was not designed for use in dangerous
atmospheres, for mixing dangerous substances or for use
under water.
If the appliance begins to slide, wipe the soft stand feet with
a damp cloth.
Contents
Safety instructions
Page
Safety instructions 6
Correct use 7
Unpacking 7
Commissioning 7
Maintenance and cleaning 7
Warranty 8
Technical data 8
In order to be able to use the appliance properly and safely,
every user must first read the operating instructions and
observe the safety instructions contained therein. Take care of
these operating instructions and keep them in a place where
they can be accessed by everyone.
The IKA lab dancer is a laboratory instrument which con-
forms to the international safety regulations for laboratory
instruments.
Caution: overfilling with liquid may cause the medium to
splash out.
You should therefore always use covered test tubes.
The unit is not suitable for manual operation. Observe all
pertinent safety instructions and directives, as well as work-
safety and accident-prevention regulations for laboratory use.