ACV; Line-level Hz (Models 037 & 033 only)
Inputs are made through the test lead terminals COM/+. Defaults at ACV Function.
Press SELECT button momentarily to select Line-level Hz function.
Note:
ACV and Line-level Hz functions are equipped with low-pass filter, and are capable of
dealing with VFD (Variable Frequency Drives) signals. It also improves reading stability
in noisy electrical environments.
DCV
Inputs are made through the test lead terminals COM/+.
Clamp-on Current Measurements: Application & Removal Of The Jaws
Press the jaw trigger and clamp the jaws around conductor(s) of only one single pole of
a circuit for load current measurements. Make sure the jaws are completely closed, or
else it will introduce measurement errors. Enclosing conductor(s) of more than one
pole of a circuit may result in differential current (like identifying leakage current)
measurements. Align the conductor(s) to the Jaws center indicators (Regular or