Fluke FLK-1507 Расширенная инструкция онлайн [20/38] 529767
![Fluke FLK-1503 Расширенная инструкция онлайн [20/38] 529766](/views2/1004883/page20/bg14.png)
1507/1503
Users Manual
12
• The primary display shows ---- until you press the
T button and a valid resistance reading is
obtained.
• The high voltage symbol (Y) along with a primary
display of >2 V warns if voltage greater than 2 V
ac or dc is present. In this condition, the test is
inhibited. Disconnect the Tester and remove
power before proceeding.
• If the Tester chirps when you press the T
button, the test is inhibited because voltage is
present at the probes.
5. Push and hold the T button to start the test. The
t icon appears on the lower portion of the display
until you release the T button. The resistance
reading appears on the primary display until a new
test is started or a different function or range is
selected.
When resistance is higher than the maximum display
range, the Tester displays the > symbol and the
maximum resistance for the range.
.
INSULATION TESTER
1507
bbw04f.eps
Figure 6. Measuring Earth-Bond Resistance
Содержание
- 1503 insulation testers users manual 1
- Insulation testers 1
- Users manual 1
- Limited warranty and limitation of liabilit 2
- Limited warranty and limitation of liability 2
- To register your product online visit register fluke com 2
- Table of contents 3
- Title page 3
- List of tables 5
- Table title page 5
- Figure title page 7
- List of figures 7
- Contacting fluke 9
- Insulation testers 9
- Introduction 9
- Safety information 10
- Xw warning 10
- Battery saver sleep mode 12
- Rotary switch positions 12
- Unsafe voltage 12
- Buttons and indicators 13
- Understanding the display 15
- Xw warning 15
- Cal err 16
- Input terminals 17
- Power up options 17
- Making measurements 18
- Xw warning 18
- Measuring earth bond resistance 19
- Measuring volts 19
- Measuring insulation resistance 21
- Measuring polarization index and dielectric absorption ratios model 1507 22
- Using the compare function model 1507 24
- Cleaning 25
- Testing the batteries 25
- Testing the fuse 26
- Xw warning 26
- Replacing the batteries and fuse 27
- Xw warning 27
- 1503 users manual 28
- General specifications 28
- Specifications 28
- Ac dc voltage measurement 29
- Accuracy 29
- Insulation testers specifications 29
- Of rdg digits 29
- Range resolution 50 hz to 400 hz 29
- V 0 v 2 3 29
- Earth bond resistance measurement 30
- Insulation specifications 30
- Model 1507 31
- En61557 specification 32
- Model 1503 32
- Insulation resistance maximum and minimum display values 34
- Insulation resistance maximum and minimum display values cont 35
- Insulation resistance maximum and minimum display values cont 36
- Insulation resistance maximum and minimum display values cont 37
- Earth bond resistance maximum display values 38
Похожие устройства
- Fluke FLK-1550C Инструкция по эксплуатации
- Fluke FLK-1550C Инструкция язык en
- Fluke FLK-1550C Data Sheet
- Fluke FLK-1550C Расширенная инструкция
- Fluke FLK-1550C-KIT Инструкция по эксплуатации
- Fluke FLK-1550C-KIT Инструкция язык en
- Fluke FLK-1550C-KIT Расширенная инструкция
- Fluke FLK-1550C-KIT Data Sheet
- Fluke FLK-1550CFC-KIT/IR Инструкция по эксплуатации
- Fluke FLK-1550CFC-KIT/IR Инструкция язык en
- Fluke FLK-1550CFC-KIT/IR Расширенная инструкция
- Fluke FLK-1550CFC-KIT/IR Data Sheet
- Fluke FLK-1550CFC/IR Инструкция по эксплуатации
- Fluke FLK-1550CFC/IR Data Sheet
- Fluke FLK-1550CFC/IR Инструкция язык en
- Fluke FLK-1550CFC/IR Расширенная инструкция
- Fluke FLK-1555 Инструкция по эксплуатации
- Fluke FLK-1555 Инструкция язык en
- Fluke FLK-1555 Расширенная инструкция
- Fluke FLK-1555 Data Sheet