Trotec OZ-ONE включая сертификат калибровки 3510006110 Инструкция к товару онлайн [7/12] 855489
![Trotec OZ-ONE включая сертификат калибровки 3510006110 Инструкция к товару онлайн [7/12] 855489](/views2/2092372/page7/bg7.png)
EN 7
ozone meter OZ-ONE
Cross sensitivities
Due to the construction of the electro chemical sensor, the
device also reacts to other gases than ozone. Besides, a high
humidity level (e.g. by blowing into the measuring probe) can
falsify the result. For this reason, always make sure that no
other gases influence the measurement and that the measuring
probe is not exposed to sources of high humidity when you
measure the ozone concentration.
Displaying DP and WBvalues
The ozone concentration is always displayed by the
measurement value display for ozone concentration(16) of the
device.
Further measured values that can be indicated are the dew
point temperatureDP and the wet-bulb temperatureWB.
Please proceed as follows to have these values displayed by the
device:
1. Repeatedly press the MODE button(8) until the desired
measured value is displayed.
µg/m
3
MODE
µg/m
3
µg/m
3
µg/m
3
Setting the HOLDfunction
1. Press the HOLD button(10).
ð The current values for the ozone concentration, relative
humidity level and ambient temperature will be held.
2. Press the HOLD button(10) again.
ð The display again indicates the current measured
values.
Setting the MAX/ MINfunction
You can view the highest or lowest value for the ozone
concentration from the current measuring interval.
To do so, please proceed as follows:
1. Briefly press the MAX/MIN button(2).
ð The MAX indication appears on the display(11) in the
measuring mode indication(14).
ð The display now indicates the highest ozone
concentration measured since the device was switched
on in the selected measuring mode.
2. Press the MAX/MIN button(2) again.
ð The MIN indication appears on the display(11) in the
measuring mode indication(14).
ð The display now indicates the lowest ozone
concentration measured since the device was switched
on in the selected measuring mode.
3. Press the MAX/MIN button(2) three times in total to return
to the normal measuring mode.
ð The display(11) again indicates the current measured
value for the ozone concentration.
Resetting the MAX/MINvalues
The MAX/MIN values of the current measuring interval can be
reset. To do so, please proceed as follows:
1. When MAX or MINindication appears on the display, press
and hold the HOLD button(10) for approx. 3seconds.
ð "CLr" will be displayed. The MAX/MIN values of the
current measuring interval will be reset and the device
automatically returns to the normal measuring mode.
Displaying the time-weighted average of the ozone
concentration
The time-weighted averageTWA indicates the average value of
the ozone concentration measured over a period of eight hours.
If the device was in operation for less than eight hours, the
average value will be calculated based on the time span since
the last switch-on.
Please proceed as follows to have the time-weighted average
valueTWA displayed by the device:
1. Repeatedly press the MAX/MIN button(2) until TWA is
indicated in the measuring mode indication(14) of the
display(11).
ð The device indicates the time-weighted averageTWA
for the measured ozone concentration.
Displaying the short-term exposure limit for the ozone
concentration
The short-term exposure limitSTEL indicates the average value
of the ozone concentration measured over a period of
15minutes. If the device was in operation for less than
15minutes, the short-term exposure limitSTEL will be
calculated based on the time span since the last switch-on.
Please proceed as follows to have the short-term exposure
limitSTEL displayed by the device:
1. Repeatedly press the MAX/MIN button(2) until STEL is
indicated in the measuring mode indication(14) of the
display(11).
ð The device indicates the short-term exposure limitSTEL
for the measured ozone concentration.
Changing the unit for the ozone concentration ppm/ μg/m
3
The device can display the ozone concentration in either ppm or
μg/m
3
. To change the units at the device, please proceed as
follows:
1. Press and hold the UNIT button(3) for approx. 3seconds to
change the unit.
ð The currently selected unit ppm or μg/m
3
will be
indicated in the unit indication for the ozone
concentration(15) of the display(11).
Похожие устройства
- Iek УКД D=37мм IP55 YUK10-37-55 Инструкция к товару
- Iek УКД D=40мм IP66 YUK10-40-66 Инструкция к товару
- Энкор 10314 Инструкция к товару
- Антиток УНП-10 БД до 10 кВ, 11 кг 315058 Инструкция
- Petzl i'd evac самоблокирующееся спусковое для высотных работ D020CA01 Инструкция
- Chint NJR2-45D 84А 45кВт 489026 Паспорт
- Chint NJR2-55D 103А 55кВт 489027 Паспорт
- Chint NJR2-7.5D 15А 7.5кВт 489019 Паспорт
- Доступная Автоматика 380в 11квт 22a modbus rtu DA40 22-A-3P3 Руководство
- Доступная Автоматика 380в 45квт 90a modbus rtu DA40 90-A-3P3 Руководство
- Innovert номинальный ток 110 а, питание управления 220в ac, rs-485 SSD553A43E Быстрый Пуск
- Innovert номинальный ток 110 а, питание управления 220в ac, rs-485 SSD553A43E Инструкция
- Innovert номинальный ток 12 а, питание управления 24в dc, rs-485 SSD222D21E Быстрый Пуск
- Innovert номинальный ток 12 а, питание управления 24в dc, rs-485 SSD222D21E Инструкция
- Innovert номинальный ток 15 а, питание управления 24в dc, rs-485 SSD752D43E Быстрый Пуск
- Innovert номинальный ток 15 а, питание управления 24в dc, rs-485 SSD752D43E Инструкция
- Innovert номинальный ток 22 а, питание управления 24в dc, rs-485 SSD113D43E Быстрый Пуск
- Innovert номинальный ток 22 а, питание управления 24в dc, rs-485 SSD113D43E Инструкция
- Innovert номинальный ток 30 а, питание управления 220в ac, rs-485 SSD153A43E Быстрый Пуск
- Innovert номинальный ток 30 а, питание управления 220в ac, rs-485 SSD153A43E Инструкция