ozone meter OZ-ONE
Cross sensitivities
Due to the construction of the electro chemical sensor, the
device also reacts to other gases than ozone. Besides, a high
humidity level (e.g. by blowing into the measuring probe) can
falsify the result. For this reason, always make sure that no
other gases influence the measurement and that the measuring
probe is not exposed to sources of high humidity when you
measure the ozone concentration.
Displaying DP and WBvalues
The ozone concentration is always displayed by the
measurement value display for ozone concentration(16) of the
device.
Further measured values that can be indicated are the dew
point temperatureDP and the wet-bulb temperatureWB.
Please proceed as follows to have these values displayed by the
device:
1. Repeatedly press the MODE button(8) until the desired
measured value is displayed.
µg/m
3
MODE
µg/m
3
µg/m
3
µg/m
3
Setting the HOLDfunction
1. Press the HOLD button(10).
ð The current values for the ozone concentration, relative
humidity level and ambient temperature will be held.
2. Press the HOLD button(10) again.
ð The display again indicates the current measured
values.
Setting the MAX/ MINfunction
You can view the highest or lowest value for the ozone
concentration from the current measuring interval.
To do so, please proceed as follows:
1. Briefly press the MAX/MIN button(2).
ð The MAX indication appears on the display(11) in the
measuring mode indication(14).
ð The display now indicates the highest ozone
concentration measured since the device was switched
on in the selected measuring mode.
2. Press the MAX/MIN button(2) again.
ð The MIN indication appears on the display(11) in the
measuring mode indication(14).
ð The display now indicates the lowest ozone
concentration measured since the device was switched
on in the selected measuring mode.
3. Press the MAX/MIN button(2) three times in total to return
to the normal measuring mode.
ð The display(11) again indicates the current measured
value for the ozone concentration.
Resetting the MAX/MINvalues
The MAX/MIN values of the current measuring interval can be
reset. To do so, please proceed as follows:
1. When MAX or MINindication appears on the display, press
and hold the HOLD button(10) for approx. 3seconds.
ð "CLr" will be displayed. The MAX/MIN values of the
current measuring interval will be reset and the device
automatically returns to the normal measuring mode.
Displaying the time-weighted average of the ozone
concentration
The time-weighted averageTWA indicates the average value of
the ozone concentration measured over a period of eight hours.
If the device was in operation for less than eight hours, the
average value will be calculated based on the time span since
the last switch-on.
Please proceed as follows to have the time-weighted average
valueTWA displayed by the device:
1. Repeatedly press the MAX/MIN button(2) until TWA is
indicated in the measuring mode indication(14) of the
display(11).
ð The device indicates the time-weighted averageTWA
for the measured ozone concentration.
Displaying the short-term exposure limit for the ozone
concentration
The short-term exposure limitSTEL indicates the average value
of the ozone concentration measured over a period of
15minutes. If the device was in operation for less than
15minutes, the short-term exposure limitSTEL will be
calculated based on the time span since the last switch-on.
Please proceed as follows to have the short-term exposure
limitSTEL displayed by the device:
1. Repeatedly press the MAX/MIN button(2) until STEL is
indicated in the measuring mode indication(14) of the
display(11).
ð The device indicates the short-term exposure limitSTEL
for the measured ozone concentration.
Changing the unit for the ozone concentration ppm/ μg/m
3
The device can display the ozone concentration in either ppm or
μg/m
3
. To change the units at the device, please proceed as
follows:
1. Press and hold the UNIT button(3) for approx. 3seconds to
change the unit.
ð The currently selected unit ppm or μg/m
3
will be
indicated in the unit indication for the ozone
concentration(15) of the display(11).