Trotec OZ-ONE включая сертификат калибровки 3510006110 Инструкция к товару онлайн [9/12] 855489

Trotec OZ-ONE включая сертификат калибровки 3510006110 Инструкция к товару онлайн [9/12] 855489
EN 9
ozone meter OZ-ONE
7. Press the HOLD button(10) to increase the value deviation.
Press the Lamp button(4) to reduce the value. You can set
a value deviation of up to ±9.9%RH.
8. Press the SETbutton(7) to save the values.
ð "SAVE" will appear on the display(11) for approx.
2seconds.
9. Press and hold the Power button(9) for approx. 3seconds.
ð The device will be switched off. The values are
adjusted.
Ozone zero adjustment
To calibrate the device, an ozone zero adjustment can be made
in an ozone-free environment. To do so, please proceed as
follows:
1. Simultaneously press and hold the UNIT(3) button and the
SET(7) button.
ð "CAL" will appear on the display(11).
2. Simultaneously press the UNIT(3) button and SET(7)
button again.
ð The calibration is started. The device counts backwards
from 120.
ð Upon successful calibration, "End" appears on the
display(11).
Info
Ensure to carry out the calibration in surroundings that
are free of other gases the sensor reacts to, e.g.
nitrogen dioxide. For more detailed information about
cross-sensitivities please refer to the chapter Errors
and faults.
De-/Activating the alarm function for the ozone limit value
Once the ozone limit value has been configured, the alarm
function can be activated or deactivated by hand.
Please proceed as follows to configure the alarm function at the
device:
1. Press the Power button(9) for approx. 1s.
ð The activated alarm indication(20) will be displayed and
the alarm function is activated.
2. Press the Power button(9) again for approx. 1s to
deactivate the alarm function.
ð The activated alarm indication(20) will no longer be
displayed and the alarm function is deactivated.
Automatic switch-off
In default setting, the automatic switch-off function is activated.
The device switches off when not in use for 15 minutes.
To deactivate automatic switch-off at the device, please
proceed as follows:
1. Switch the device off by pressing the Power button(9) and
holding it for approx.3s.
2. Simultaneously press the Power button(9) and the HOLD
button(10) until the message "n" is displayed.
ð Automatic switch-off is permanently deactivated.
Info
If you switch off the device manually, the automatic
switch-off setting is reset, i.e. automatic switch-off is
activated again when the device is switched back on.
Switch-off
If automatic switch-off is activated, the device switches off after
15minutes of non-use.
1. Press and hold the Power button(9) for approx. 3seconds.
ð The device will be switched off.