Ada COSMO 40 А00376 [15/60] Cosmo 40 1

Ada COSMO 40 А00376 [15/60] Cosmo 40 1
COSMO 40
15
INFO CAUSE REMEDY
204
Calculation overflow Repeat procedure
205
Out of measuring range Choose measuring distance within measuring range
252
Temperature too high Cool down instrument
253
Temperature too low Warm up instrument
255
Receiver signal too weak Use target plate
MESSAGE CODES
All message codes are displayed with either “Info” or symbol “telephone receiver” (Error). Following
mistakes can be corrected.