Warranty
Respironics New Jersey, Inc. warrants the Family compressor to be free from defects in materials and
workmanship under normal use and operation for a period of 3 years from date of purchase from Respironics
New Jersey, Inc. The warranty is limited to repair or replacement, at Respironics New Jersey, Inc.’s sole
option, of any such component or equipment claimed to be defective when claim is shown to be bona fide by
evaluation by Respironics New Jersey, Inc. This warranty does not extend to any components or equipment
subjected to misuse, improper operation, accidental damage, or unauthorized repairs, and is not extended to
charges of, or for, labor repairs. All items returned must be properly packaged and shipped, prepaid, by the
distributor servicing the unit. Respironics New Jersey, Inc. shall not be liable to purchaser or others for loss
of use of equipment or for indirect, or incidental or consequential damages that might arise.
Device symbols
l ON (power)
O OFF (power)
Type B
Separate collection
ATTENTION: Read the instructions carefully
Alternating current
Class II
~
Éléments de l’appareil Family
1. Compartiment du compresseur
2. Interrupteur de marche/arrêt (ON / OFF)
3. Cordon d’alimentation
4. Sortie d’air
5. Logement du filtre
6. Logement du nébuliseur
7. Ouvertures d’aération
8. Poignée
Accessoires standard
9. Housse de transport (uniquement pour le modèle Silver)
10. Nébuliseur Jet Basic
11. Embout buccal coudé
12. Embout nasal
13. Tube à air
14. Masque pour adultes
15. 5 Filtres de rechange
EN
E
FR
Technical specifications
Model Family
Electricity supply 230V/50Hz (#90479, 90543)
Nebulization rate 0.28 ml/min
Particle size MMD 4.99 µm
Compressor max. pressure 230 kPa (2.30 bar)
Weight 1.55 kg
Size (H) 167 x (L) 145 x (W) 97 mm
Noise level 55 dBA
Class II device (device with specific protection against electrical hazards).
Type B device (device with specific protection against electrical hazards).
Device should not be used around flammable substances, or in the presence of a flammable anesthetic
mixture with air, oxygen or nitrous oxide.
Device not suitable for use in anesthetic or lung ventilation systems.
Reference to standards
Electric safety standards EN 60601-1
Electromagnetic compatibility according to EN 60601-1-2
Ambient conditions
Storage conditions
Temperature MIN -25 °C MAX 70 °C (-13 °F 158 °F)
Humidity MIN 10% RH MAX 95% RH
Operating conditions
Temperature MIN 10 °C MAX 40 °C (50 °F –104 °F)
Humidity MIN 10% RH MAX 95% RH
Altitude From 0 to 2000m above sea level
Atmospheric Pressure 700 hPa 1060 hPa