9 Tips and assistance
45
9 Tips and assistance
9.1 Questions and answers
Question Possible cause / solution
Error! Memory full! is
displayed.
Insufficient memory available: Transfer images
to the PC or delete.
Error! Permissible
instrument temperature
exceeded! is displayed.
Switch off the camera, allow the instrument to
cool down and observe the permissible
ambient temperature.
~ is displayed before a value.
Value is outside the measuring range:
extended display range with no guarantee of
accuracy.
--- or +++ is displayed
instead of a value.
Value is outside the measuring range and the
extended display range.
xxx is displayed instead of a
value.
Value cannot be calculated: check parameter
settings for plausibility.
A
utomatic zeroing (audible
"click" and brief image freeze)
is carried out very frequently.
Camera is still in its warm-up period (takes
approx. 90 seconds): Wait until the warm-up
period has passed.
If we have not been able to answer your question, please contact your dealer or
Testo Customer Service. You will find contact details on the back of this
document or on the website www.testo.com/service-contact.