Fluke 971 [5/5] Change 4
Manual Supplement 971 Users
4 3/19
Change #4, 396, 684
On page 7, following the Cleaning section
add:
Humidity Sensor Reconditioning
If unit is found to be out of specification for
humidity, the sensor could be contaminated by
chemical vapors or offset due to long exposure
in a dry environment. This condition is typical
of polymer type humidity sensors. If
contaminated, the sensor will need to be
baked, and rehydrated.
Bake: >50 °C at <5 % RH for 24 hours
Hydration: 20 °C to 30 °C @ 75 % RH for
48 hours
If offset is only due to a long exposure in a dry
environment, sensor will only need to be
rehydrated.
Hydration: 20 °C to 30 °C @ 75 % RH for
48 hours