Fluke 754 [2/6] Change 1

Fluke 754 [2/6] Change 1
753/754 Users Manual Supplement
1/17 1
Change #1, 64190, 64295, 544
On page 5, remove the ¤ from the symbols table, add the following to the Symbols table:
à Conforms to relevant South Korean EMC Standards.
» Consult user documentation.
Replace the ; row with:
¡ Conforms to relevant Australian Safety and EMC standards.
On page 106, under General Specifications, replace Power with:
Power.......................................................................... Internal battery pack: Lithium Ion, 7.2 V dc, 30 Wh
External Power Supply: 15 Vdc, 2 A
On page 107, replace the Standards and Agency Approval Specifications with:
Standards and Agency Approval Specifications
Intrusion Protection ................................................... IEC 60529-1, IP 52
Safety
General .................................................................... IEC 61010-1, Pollution degree 2
Measurement ........................................................... IEC 61010-30: CAT II 300 V