719 Users Manual Supplement
3/09 1
Change #1, 48042 and 49682
On page 26, replace the Pressure Sensor Input and the DC mA Measure and Source tables
with the following:
Pressure Sensor Input
Model Range Resolution Accuracy
[1]
Max Non-destructive Pressure
30G -12 to 30.0 psi 0.001
0.025 % of Range
(6 month calibration)
0.035 % of Range
(1 year calibration)
60 psi
100G -12 to 100.0 psi 0.01 200 psi
Temperature coefficient: 0.01 % of range per C for temperature ranges -10 C to 18 C and 28 C to 55 C
[1] In an RF field of 1V/m to 3V/m between the frequencies of 1.6 GHz to 2.0 GHz, add 120 counts to specified
accuracy.