Fluke 721 [2/2] Change 1
![Fluke 721 [2/2] Change 1](/views2/1436149/page2/bg2.png)
721 Calibration Manual Supplement
12/17 1
Change #1, 532
On page 6, replace Table 2 with:
Equipment Minimum Specifications Recommended Model
Process Calibrator With Simulate Transmitter Mode Fluke 754
Precision DMM Accuracy: 0.003 % rdg ±0.5 µA Fluke 8508A
On page 11, replace the mA Loop Performance Check section with:
mA Loop Verification
1. Set the Process Calibrator to source mA in the Simulate Transmitter mode.
2. Set the Precision DMM to measure mA dc.
3. Connect the Process Calibrator mA red jack to the product black jack.
4. Connect the Product red jack to the Precision DMM Input LO.
5. Connect the Precision DMM Input A to the Process Calibrator mA black jack.
6. Push F2 repeatedly until mA Loop is shown in the third row of the Product display.
7. Use the Process Calibrator to supply the currents in the table below. Calculate the limits,
and verify whether the Product is within the limits.
Applied from
Process
Calibrator
Reference Reading
from Precision DMM
Upper Limit
(Reference Reading +
0.015 % of Reference
Reading + 0.002 mA)
Lower Limit
(Reference Reading -
0.015 % of Reference
Reading - 0.002 mA)
4 mA
12 mA
20 mA
24 mA
On page 18, replace item number 12 with:
Fluke-721-2003, Holster, Yellow 1
4497298
Содержание
- Manual supplement 1
- Calibration manual supplement 2
- Change 1 2
- Ma loop verification 2
- On page 11 replace the ma loop performance check section with 2
- On page 18 replace item number 12 with 2
- On page 6 replace table 2 with 2
- Set the process calibrator to source ma in the simulate transmitter mode 2 set the precision dmm to measure ma dc 3 connect the process calibrator ma red jack to the product black jack 4 connect the product red jack to the precision dmm input lo 5 connect the precision dmm input a to the process calibrator ma black jack 6 push f2 repeatedly until ma loop is shown in the third row of the product display 7 use the process calibrator to supply the currents in the table below calculate the limits and verify whether the product is within the limits 2
Похожие устройства
- Fluke 721 Положение об энергозависимости памяти
- Fluke 721 Паспорт безопасности
- Fluke 721 Спецификация
- Fluke 721 Руководство по калибровке
- Fluke 721 Краткий справочник
- Fluke 721 Pressure Calibration APPLICATIONS AND SOLUTIONS
- Fluke 721 Руководство по эксплуатации
- Fluke 700G Положение об энергозависимости памяти
- Fluke 700G Data Sheet
- Fluke 700G Приложение к руководству по калибровке
- Fluke 700G Паспорт безопасности
- Fluke 700G Руководство по калибровке
- Fluke 700G Приложение к Руководству по эксплуатации
- Fluke 700G Premium Pressure Accessory Kit _ Краткий справочник
- Fluke 700G Руководство по эксплуатации
- Fluke 700G TRACK _ Руководство по эксплуатации
- Fluke 750P00 Положение об энергозависимости памяти
- Fluke 750P00 Приложение к Инструкции
- Fluke 750P00 Инструкция
- Fluke 750P01 Положение об энергозависимости памяти