Fluke P3800 [9/32] Chapter 1
![Fluke P3800 [9/32] Chapter 1](/views2/1437093/page9/bg9.png)
1-1
Chapter 1
General Information
Introduction
This manual contains operation and routine and preventive maintenance instructions for
the Model P3830, P3840, and P3860 High Pressure Hydraulic Deadweight Tester (DWT)
manufactured by Fluke. This section of the manual provides general information about
the DWT.
The P3800 Series of Deadweight Testers (DWT) provides a convenient means of testing
high pressure instruments for calibration accuracy.
The design incorporates the “Piston Gauge” principle, in which an applied pressure
within the system, balances a known mass applied to a piston of known effective area. i.e.
pistonofareaEffective
pistontoappliedmassTotal
PRESSURE =
The unit comprises a base plate onto which is mounted a high pressure manifold, a
piston/cylinder unit (PCU), a test station (to mount the equipment under test), a fluid
reservoir, a ram screw pressure generating system, a pressure intensifier (which increases
the pressure generated by a factor of 7:1), and a safety relief valve (which protects both
the low and high pressure systems from accidental overpressure). The whole assembly is
contained within a strong GRP housing.
The unit is equipped with four adjustable feet to enable correct leveling. This ensures that
the PCU is in the vertical plane, which is essential for both accurate and reliable
performance.
The hydraulic fluid supplied with the unit has been specially formulated to remain fluid at
high pressures and to ensure no corrosion of internal parts. In addition the fluid will not
oxidize in contact with air or emulsify or mix with water.
How to Contact Fluke
To order accessories, receive operating assistance, or get the location of the nearest Fluke
distributor or Service Center, call:
• Technical Support USA: 1-800-99-FLUKE (1-800-993-5853)
• Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
• Canada: 1-800-36-FLUKE (1-800-363-5853)
• Europe: +31-402-675-200
• China: +86-400-810-3435
• Japan: +81-3-3434-0181
• Singapore: +65-738-5655
• Anywhere in the world: +1-425-446-5500
Содержание
- P3800 series 1
- P3800 series users manual 1
- Warranty 2
- Table of contents 3
- List of tables 5
- List of figures 7
- Chapter 1 9
- General information 9
- How to contact fluke 9
- Introduction 9
- Piston of area effective 9
- Piston to applied mass total pressure 9
- Safety information 10
- Symbols used in this manual 10
- W warning 10
- General information 11
- Hydraulic circuit schematic 11
- Chapter 2 13
- Fluid filling 13
- Location 13
- Preparation 13
- Chapter 3 15
- Connections 15
- Operation 15
- P3800 series 16
- Procedure for removing trapped air from system 16
- Procedure for priming the pressure intensifier 18
- Procedure for generating system pressure 19
- Generating calibration pressure 20
- Chapter 4 23
- Introduction 23
- Maintenance 23
- Pcu removal 23
- W warning 23
- Cleaning 24
- Replacement 25
- Chapter 5 27
- Do s and don ts 27
- Don ts 27
- Introduction 27
- Recalibration 27
- Chapter 6 29
- If the dwt is located at a position where the values of the gravitational acceleration and temperature of the operation are the same as those values specified for calibration then the actual pressure is fully corrected if the dwt is used under different conditions then further correction is necessary the reference level is specified on the certificate 29
- Introduction 29
- Pressure correction is required for high accuracy work and is due to the effects of pressure on the pcu assembly during operation by reference to the certificate of calibration provided the actual pressure in the system can be obtained all values relate to the environmental conditions stated on the certificate 29
- Pressure corrections 29
- The pressure in the system when the carrier is loaded with incremental weights is given together with the pressure in the system when the carrier is loaded with major and incremental weights from these values assuming effects are linear the incremental weight corrections over the range may be calculated 29
- The pressure in the system when the carrier is loaded with major weights can be read directly from the second column of the certificate 29
- Temperature and gravity corrections 30
- Explanation of nomogram a straight line passing through the known values of altitude h and latitude l of the site of the dwt when extended to scale g will indicate the approximate value of g 31
- H 16000 31
Похожие устройства
- Fluke P3800 Pressurements PressCal _ Руководство по эксплуатации
- Fluke P3200 Pressurements PressCal _ Руководство по эксплуатации
- Fluke 714B Положение об энергозависимости памяти
- Fluke 714B Приложение к паспорту безопасности
- Fluke 714B Приложение к Руководству по эксплуатации
- Fluke 714B Паспорт безопасности
- Fluke 714B Спецификация
- Fluke 714B Краткий справочник
- Fluke 714B Руководство по эксплуатации
- Fluke 714B Руководство по калибровке
- Fluke 714C Дополнение к краткому справочному руководству
- Fluke 714C Приложение к Руководству по эксплуатации
- Fluke 714C Руководство по эксплуатации
- Fluke 714C Краткий справочник
- Fluke 712B Положение об энергозависимости памяти
- Fluke 712B Приложение к Руководству по эксплуатации
- Fluke 712B Паспорт безопасности
- Fluke 712B Приложение к паспорту безопасности
- Fluke 712B Руководство по эксплуатации
- Fluke 712B Краткий справочник