Ada Metrix 60 А00105 [33/40] Metrix 60

Ada Metrix 60 А00105 [33/40] Metrix 60
Metrix 60
MEASUREMENT FOUNDATION
33
Troubleshooting
All displayed codes come with symbol Err (Mistake).
Measuring conditions
Measuring range
The range is limited to 60 m.
At night or in the dark and if the target is in the shadow the measuring range without target plate is increased. Use the
target plate to increase the measurement range during daylight or if the target has poor reection properties.
Code Cause Corrective measure
204 Calculation error Repeat procedure
208 Received signal too weak,
measurement time too long.
Distance >50 m.
Use target plate
209 Received signal too strong Use target plate
252 Temperature too high Cool down the instrument
253 Temperature too low Warm up the instrument
255 Hardware error Switch on/off the device several
times. If the symbol still appears,
please contact the dealer for
assistance.