Fluke 353 [24/26] Ohms measurement 355 only
![Fluke 353 [24/26] Ohms measurement 355 only](/views2/1005189/page24/bg18.png)
353/355
Users Manual
16
Voltage Measurement (355 only) 10 Hz to 100 Hz
600 and 1000 V ranges have 10 % over range to 660 and 1100 V respectively.
Range Resolution Accuracy
Trigger Level for Hz
Filter OFF
Trigger Level for Hz
Filter ON
4 V 1 mV 1 % rdg + 10 digits 0.050 V 0.050 V
40 V 10 mV 1 % rdg + 5 digits 0.25 V 0.25 V
400 V 100 mV 1 % rdg + 5 digits 6 V 6 V
600 V AC RMS 1 V 1 % rdg + 5 digits 6 V 6 V
1000 V DC 1 V 1 % rdg + 5 digits N/A N/A
Voltage Measurement (355 only) 100.1 Hz to 1 kHz
600 and 1000 V ranges have 10 % over range to 660 and 1100 V
Range Resolution Accuracy
4 V 1 mV 3 % rdg + 10 digits
40 V 10 mV 3 % rdg + 5 digits
400 V 100 mV 3 % rdg + 5 digits
600 V AC RMS 1 V 3 % rdg + 5 digits
Ohms Measurement (355 only)
Range Resolution Accuracy
400 Ω 0.1 Ω 1.5 % + 5 digits
4 k Ω 1 Ω 1.5 % + 5 digits
40 k Ω 10 Ω 1.5 % + 5 digits
400 k Ω 100 Ω 1.5 % + 5 digits
Содержание
- 355 users manual 1
- Clamp meter 1
- Warranty 2
- Table of contents 3
- Title page 3
- List of tables 5
- Table title page 5
- Figure title page 7
- List of figures 7
- Clamp meter 9
- Contacting fluke 9
- Introduction 9
- Safety information 10
- Xw warnings read first 10
- Features 13
- Display 15
- Maxminavglive 15
- Measuring ac or dc current 16
- Using the meter 16
- Xw warning 16
- Measuring inrush current 17
- Measuring ac and dc voltage 355 only 18
- Clamp meter using the meter 19
- Fbq06 eps 19
- Figure 6 dc voltage measurement 19
- Testing continuity 355 only 20
- Xw warning 20
- Measuring resistance 355 only 21
- Xw warning 21
- Changing the batteries 22
- Cleaning the meter 22
- Maintenance 22
- User replaceable parts 22
- Xw warning 22
- Current measurement 10 hz to 100hz 23
- Current measurement 100 hz to 1 khz 23
- Electrical specifications 23
- Specifications 23
- Ohms measurement 355 only 24
- Voltage measurement 355 only 10 hz to 100 hz 24
- Voltage measurement 355 only 100 hz to 1 khz 24
- Continuity beeper 355 25
- Frequency measurement 25
- General specifications 25
- Environmental specifications 26
- Standards and agency approval specifications 26
Похожие устройства
- LG F1480TDS Инструкция по эксплуатации
- Ada 5D Crosspoint А00135 Инструкция по эксплуатации
- Supra SWD-703B Инструкция по эксплуатации
- DBX PROVOCAL Инструкция по эксплуатации
- LG F1073TD Инструкция по эксплуатации
- LG MS-2342W Инструкция по эксплуатации
- Fluke 355 Инструкция по эксплуатации
- Boif AL120 В00100 Инструкция по эксплуатации
- Supra SWD-605 Инструкция по эксплуатации
- LG MS-2342A Инструкция по эксплуатации
- Fluke 373 Инструкция по эксплуатации
- Bosch WLX 24463 OE Инструкция по эксплуатации
- DBX 1215 Инструкция по эксплуатации
- Geo-Fennel Ecoline EL601 D1200 Инструкция по эксплуатации
- Supra SDM-3150 Инструкция по эксплуатации
- Belkin F7D1301ru Инструкция по эксплуатации
- Fluke 374 Инструкция по эксплуатации
- Bosch WLX 20463 OE Инструкция по эксплуатации
- DBX 1231 Инструкция по эксплуатации
- LG MS-2324B Инструкция по эксплуатации