IKA C-MAG HS 7 digital 0003487000 [5/6] Custom safety temperature

IKA Plate (RCT digital) 0025004601 [5/6] Custom safety temperature
Adjust your individual safety temperature
device as follows:
A small tool is provided as an accessory for all IKA laboratory devices with Integrated
Safety which you can manually adjust the safety temperature for every laboratory
procedure.
The screw opening provided for this is usually next to the On/Off switch on the
device. Insert the tool through the opening and turn it until the desired maximum
temperature is reached.
Only by manual adjustment using a tool can the total independence
of the safety circuit from the normal control circuit required in DIN
EN IEC 61010-2-010 be achieved.
Further detailed information can be taken from the operating manual for your
specific device.
Custom safety temperature
Since every sample is different, it is possible to individually set the maximum
temperature allowed to ensure its safety. Integrated Safety from IKA ensures that the
respective flashpoint of the medium used is not reached, independently of any device
defects, glass breakages, or physical effects such as thermal distortion.