3. Raise the hand and attach the Smart indicator AGU STI2
just below the armpit. Be sure to place the Smart indicator
as close to the axillary as possible and place it the same
way as on the picture is shown.
WARNING
Ideal temperature monitoring points are body
cavities close to major blood vessels: the mouth,
rectum, under-arm (axillary) and eardrum. Tem-
perature measured from other position can’t be
considered as core body temperature of human
being which might cause wrong judgement.
USING A SMART TEMPERATURE INDICATOR AGU STI2
WITH A ONE-SIDED STICKER
To use the one-sided sticker will further avoid movement of
the Smart temperature indicator AGU STI2.
1. Remove the protective layer of the one-sided sticker.
2. Use the sticker to x the Smart indicator which already
attached on the under-arm.
3. Make sure that the button hole in the one-sided sticker
matches the button on the device case.