Do not exceed the motor speed recommended by the handpiece and bur manufacturer Never lubricate the motor Oil could cause overheating and damage to the motor Do not use high acid water or sterilizing solutions to wipe immerse or clean the product The motor and motor cord cannot be cleaned and disinfected with a Thermo Disinfector Do not sterilize the motor and motor cord Perform regular function and maintenance checks If the product is not used for a long period check it is functioning correctly before using on a patient Air Requirements The air should be dry free from contamination and oil Use a compressor with a dry air system Install an air filter if necessary Blow out the lines before installation Do not use this motor under stress for long time as this will cause overheating Do not pull the Motor excessively as this may loosen or disconnect the motor cord and cause malfunction Surface temperature shall be more than 60 C without using a coolant air To avoid this occurrence be sure to use sufficient coolant air When using the micromotor for more than 2 minutes continuously it will become hot The maximum surface temperature might reach 45 C In such cases stop using the micromotor until the motor has cooled down sufficiently To avoid clinical downtime it is recommended that a spare be kept on hand in case of a breakdown during surgery Avoid continual eye contact with the LED light M40 XS This product is rated Medical Electrical equipment EMC Electromagnetic compatibility is described in the documentation included Installation and use of this product requires special precautions regarding EMC according to the EMC information