Install the instrument in a place where the ambient temperature is in range of 10~70°C. Note that the instrument
operation may become unstable if it is operated in an ambient temperature exceeding the range of 0~40°C
2.3. Power Supply
AC Input voltage: 110V~120V/220~240V ±10%, 50/60Hz.
Connect the protective grounding conductor of the AC power cord to an earth ground to avoid electrical shock.
2.4. Fuse
Fuse type: 110~120V: T2A /250V, or 220~240V: T1A/250V.
Make sure the correct type of fuse is installed before power up.
Replace the AC fuse with the same type and rating as the original fuse.
Disconnect the power cord before fuse replacement.
Make sure the cause of fuse blowout is fixed before fuse replacement.
2.5. Maintenance Inspection
Inspect the instrument at regular intervals so that it maintains its initial performance for a long time.
Check the input power cord for damage of the vinyl cover and overheating of the plug and cord stopper. Check the
terminal screws and binding posts for loosening.
Remove dust from the inside of the casing and ventilation holes of the cover by using a compressed air of the exhaust
air of a vacuum cleaner.
2.6. Cleaning
Before cleaning, disconnect the AC mains.