B&K PRECISION BK5492CGPIB [5/10] Specifications 5492c 5492cgpib

B&K PRECISION BK5493CGPIB [5/10] Specifications 5492c 5492cgpib
Digital Multimeters
5490C Series
5
www.bkprecision.com
Specifications - 5492C / 5492CGPIB
Range
(3)
Resolution Full Scale Reading
100 mV 1 µV 119.999 mV
1 V 10 µV 1.19999 V
10 V 100 µV 11.9999 V
100 V 1 mV 119.999 V
750 V
(4)
10 mV 787.50 V
(1) Valid for sine wave input > 0.3% of range and greater than 1 mVrms.
(2) Three filters are available for low-frequency performance: 3 Hz, 20 Hz, and 200 Hz. No additional errors will occur for frequencies above these filter settings.
(3) 20% overrange on all ranges except 750 V range can only test 5% overrange.
(4) 750 V range is limited to 8 x 10
7
Volt-Hz.
Range
(7)
Burden Voltage Resolution
Full Scale
Reading
Accuracy (1 year)
3 Hz to 5 Hz 5 Hz to 20 Hz 20 Hz to 10 kHz
100 µA < 0.011 V 1 nA 119.999 µA 1.00 + 0.03 0.5 + 0.03 0.20 + 0.04
1 mA < 0.11 V 10 nA 1.19999 mA 1.00 + 0.03 0.5 + 0.03 0.20 + 0.04
10 mA < 0.05 V 0.1 µA 11.9999 mA 1.00 + 0.03 0.5 + 0.03 0.20 + 0.04
100 mA < 0.5 V 1 µA 119.999 mA 1.00 + 0.03 0.5 + 0.03 0.20 + 0.04
1 A < 0.7 V 10 µA 1.19999 A 1.00 + 0.03 0.5 + 0.03 0.20 + 0.04
3 A < 2.0 V 10 µA 3.1500 A 1.00 + 0.03 0.5 + 0.03 0.25 + 0.04
10 A
(8)
< 0.5 V 100 µA 11.9999 A 1.00 + 0.03 0.5 + 0.03 0.30 + 0.04
True RMS AC Current
(5)(6)
(5) Three filters are available for low-frequency performance: 3 Hz, 20 Hz, and 200 Hz. No additional errors will occur for frequencies above these filter settings.
(6) Valid for sinewave input > 1% of range and greater than 10 µA AC.
(7) 20% overrange on all ranges except 3 A range can only test 5% overrange.
(8) 10 A range available on front panel only.
Range Test Current Accuracy (1 year) Temperature Coefficient / °C
1 kΩ 1 mA 0.030 + 0.004 0.0010 + 0.0020
Continuity
(9) Applicable to the voltage measured at the input terminal. A 1 mA test current is typical. Changes in the current source will
cause variations in voltage drop across the diode junction.
Range Test Current Accuracy (1 year) Temperature Coefficient / °C
5 V 1 mA 0.010 + 0.005
(1)
0.0010 + 0.0020
Diode
(9)
Type Accuracy
Pt100 (DIN / IEC 751) Probe accuracy + 0.05 °C
5 kΩ thermistor Probe accuracy + 0.10 °C
Temperature
Frequency Accuracy (1 Year)
3 Hz to 5 Hz 1.00 + 0.03
5 Hz to 20 Hz 0.38 + 0.03
20 Hz to 20 kHz 0.10 + 0.03
20 kHz to 50 kHz 0.15 + 0.05
50 kHz to 100 kHz 0.63 + 0.08
100 kHz to 300 kHz 4.00 + 0.50
True RMS AC Voltage
(1)(2)
100 mV, 1 V, 10 V, 100 V, and 750 V ranges