CHAUVIN ARNOUX CA-MN77 [12/16] Oscilloscope clamp for ac current model mn60

CHAUVIN ARNOUX CA-MN77 [12/16] Oscilloscope clamp for ac current model mn60
Serie MN
Non-contractual document
906131102B - Ed 1
2.08 (3/4)
MN series
Oscilloscope clamp for AC current
Model MN60
(insulated AC current probe)
FREQUENCY RESPONSE (CONT.)
200 A calibre 20 A calibre
Frequency in Hz
Error on measurement in %
-10.00
-8.00
-6.00
-4.00
-2.00
0.00
2.00
10.00 100.00 1000.00 10000.00 100000.00
E % at 100 A
E % at 200 A
Frequency in Hz
Error on measurement in %
-30.00
-25.00
-20.00
-15.00
-10.00
-5.00
0.00
5.00
10.00 100.00 1000.00 10000.00 100000.00
E % at 10 A
E % at 20 A
Frequency in Hz
Phase in °
-25.00
-20.00
-15.00
-10.00
-5.00
0.00
5.00
10.00
15.00
10.00 100.00 1000.00 10000.00 100000.00
Phase at 100 A
Phase at 200 A
Frequency in Hz
Phase in °
-40.00
-30.00
-20.00
-10.00
0.00
10.00
20.00
30.00
40.00
10.00 100.00 1000.00 10000.00 100000.00
Phase at 10 A
Phase at 20 A
RESPONSE TO A SQUARE SIGNAL
200 A calibre 20 A calibre
10 A at 10 Hz
10 A at 100 Hz