380950-EU-EN-V1.1 7/12
5
Diode Test
1. Turn the rotary switch to the
•))) CAP position.
2. Insert the black test lead banana plug into the negative (COM) jack
Insert the red test lead banana plug into the positive (VHz jack.
3. Push the mode button to indicate on the display.
4. Touch the test probes to the diode under test. Typically for a normal diode, forward
voltage will indicate 0.4V to 0.7V. Reverse voltage will indicate “OL”. Shorted
devices will indicate near 0V and an open device will indicate “OL” in both polarities.
Capacitance Measurements
Warning: To avoid electrical shock, disconnect power to the unit under test and discharge all
capacitors before taking any capacitance measurements. Remove the batteries and unplug the
line cords.
1. Set the function switch to the •))) CAP position.
2. Push the mode button to indicate nF on the display.
3. Insert the black lead banana plug into the negative (COM) jack
Insert the red test lead banana plug into the positive (VHz) jack.
4. Press the ZERO key to null the meter display.
5. Touch the test probe tips to the capacitor you wish to check.
6. Read the capacitance value on the display.
Frequency or % Duty Cycle Measurements
1. Turn the rotary switch to the Hz % position.
2. Insert the black test lead banana plug into the negative (COM) jack and the red test
lead banana plug into the positive (VHz jack.
3. Select Hz or % with the HZ/% button.
4. Touch the test probe tips to the circuit under test.
5. Read the frequency on the display.
Analog Signal Output
1. Turn the rotary switch to the DCA or ACA ranges.
2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead banana
plug into the positive (VHz jack.
3. Connect the test leads to a multimeter, oscilloscope or chart recorder inputs.
4. Press the trigger to open the jaw. Fully enclose the conductor to be measured.
5. The analog voltage signal is output to the measuring device.
Note: When measuring DCA, the output signal is DCV. When measuring ACA, the output signal is
both ACV and DCV.
Содержание
Похожие устройства
- EXTECH EX380976-K Инструкция по эксплуатации
- EXTECH EX380976-K Инструкция язык en
- EXTECH EX380976-K Data Sheet
- EXTECH EX613 Инструкция по эксплуатации
- EXTECH EX613 Data Sheet
- EXTECH EX613 Расширенная инструкция
- EXTECH EX613 Инструкция язык en
- EXTECH EX612 Инструкция по эксплуатации
- EXTECH EX612 Data Sheet
- EXTECH EX612 Инструкция язык en
- EXTECH EX623 Инструкция по эксплуатации
- EXTECH EX623 Data Sheet
- EXTECH EX623 Расширенная инструкция
- EXTECH EX623 Инструкция язык en
- EXTECH EX655 Инструкция по эксплуатации
- EXTECH EX655 Data Sheet
- EXTECH EX655 Инструкция язык en
- EXTECH EX655 Расширенная инструкция
- EXTECH EX830 Инструкция по эксплуатации
- EXTECH EX830 Data Sheet