380950-EU-EN-V1.1 7/12
8
Jaw size 12.7mm (0.5") approx.
Display 4000 count LCD
Continuity Audible tone < 150 approx.
Diode Test Open circuit voltage < 1.5VDC; Test current <1mA (typical)
AC V bandwidth 50Hz to 400Hz
AC A bandwidth 50/60Hz
Low battery indication “ ” is displayed
Overrange indication “OL” is displayed
Auto Power OFF After 25 minutes
Measurement rate 2 per second, nominal
Input Impedance 7.8M (V DC and V AC)
Operating Temperature -10ºC to 50ºC (14ºF to 122ºF)
Storage Temperature -30
o
C to 60
o
C (-22
o
F to 140
o
F)
Operating Humidity Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at 45ºC (113ºF)
Storage Humidity <80%
Operating Altitude 2000meters (6560ft) operating
Batteries (2) 1.5V AAA batteries
Weight 200g (0.44lb)
Size 200x50x35mm (7.87” x 1.97” x 1.38”)
Safety For indoor use and in accordance with the requirements for double
insulation to IEC1010-1 (1995): EN61010-1 (1995) Overvoltage
Category III, Pollution Degree 2.
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORIES
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which
measures are taken to limit the transient overvoltages to an appropriate low level.
Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from
the fixed installation.
Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.
Note – Examples include switches in the fixed installation and some equipment for industrial use
with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.
Note – Examples include electricity meters and primary over-current protection equipment