Environmental
Pollution degree .......................2
Protection degree .....................IP54 (T90)
IP64 (T110/T130/
T150)
Operating Temperature ............-15 °C to +45 °C
Storage Temperature ................-20 °C to +60 °C
Humidity ...................................85 % RH maximum
Altitude .....................................2000 m
Vibration ...................................refer to EN61243-3
Safety EN61243-3:2010
Transporting goods...................VBG 1, § 35
Overvoltage protection .............690 V ac/dc
Measurement category
T90......................................CAT II 690 V
CAT III 600 V
T110/T130/T150 .................CAT III 690 V
CAT IV 600 V
Power supply ...............................2 x 1.5 V Micro /
LR03 / AAA
Power consumption ....................50 mA maximum /
~250 mW
Language support .......................Czech, Dutch,
English, Finnish,
French, German,
Italian, Norwegian,
Polish, Portuguese,
Romanian,
Russian, Spanish,
Swedish, Turkish
LIMITED WARRANTY & LIMITATION OF
LIABILITY
This Fluke product will be free from defects in material and
workmanship for two years from the date of purchase. This
warranty does not cover fuses, disposable batteries, or damage
from accident, neglect, misuse, alteration, contamination, or
abnormal conditions of operation or handling. Resellers are not
authorized to extend any other warranty on Fluke’s behalf. To
obtain service during the warranty period, contact your nearest
Fluke authorized service center to obtain return authorization
information, then send your defective Tester to that Service
Center with a description of the problem. Replace depleted
batteries immediately to avoid Tester damage from battery
leakage.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT
LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM
ANY CAUSE OR THEORY. Since some states or countries do
not allow the exclusion or limitation of an implied warranty or of
incidental or consequential damages, this limitation of liability
may not apply to you.
Fluke Corporation Fluke Europe B.V
P.O. Box 9090 P.O. Box 1186
Everett WA 98206‑9090 5602 B.D.
Eindhoven
The Netherlands
11/99
Specications
Model
T90 T110 T130 T150
LEDs
Voltage range 12 V to 690 V ac/dc
• • • •
Resolution
±12 V, 24 V, 50 V, 120 V,
230 V, 400 V, 690 V
• • • •
Tolerance
Complies with
EN 61243-3:2010
• • • •
Frequency range 0 / 40 Hz to 400 Hz
• • • •
Response time ≤0.5 second
• • • •
Auto power on ≥12 V ac/dc
• • • •
LCD
Voltage range 6 V to 690 V ac/dc
• •
Resolution ±1 V
• •
Tolerance ±(3 % rdg + 5 digits)
• •
Frequency range 0 / 40 Hz to 400 Hz
• •
Response time ≤1 second
• •
Auto power on ≥6 V ac/dc
• •
Voltage detection Automatic
• • • •
Polarity detection Full range
• • • •
Range detection Automatic
• • • •
Internal basic load impedance
Peak current
Maximum 3.5 mA at
690 V 200 kΩ / Is <3.5 mA
(no RCD tripping)
• • • •
Operation time
Duration Time =
30 seconds
• • • •
Recovery time
Recovery Time =
240 seconds
• • • •
Switchable Load ~7 kΩ
• • •
Peak current Is (load) = 150 mA
• • •
RCD tripping I~30 mA @ 230 V
• • •
Continuity Test 0 to 400 kΩ
• • • •
Accuracy nominal resistance +50 %
• • • •
Test current ≤5 µA
• • • •
Single-pole Phase Test 100 V ac to 690 V ac
• • • •
Frequency range
40 Hz to 60 Hz
•
50 Hz to 400 Hz
• • •
Rotary Field Indication
• • •
Voltage range (LEDs)
100 V to 690 V
(phase to earth)
• • •
Frequency range 50 Hz to 60 Hz
• • •
Resistance Measurement 0 Ω to 1999 Ω
•
Resolution 1 Ω
•
Tolerance
±(5 % rdg +10 digits) @
20 °C
•
Temperature coefcient ±5 digits / 10 K
•
Test current ≤30 µA
•
Size in mm (HxWxL) 245x64x28 255x78x35
Weight in kg (includes batteries) 0.18 0.27
How to Clean
Before you clean the Tester, remove it from all
measurement circuits.
Caution
To prevent damage, do not use abrasives or
solvents on the Tester.
Clean the case with a moist cloth and weak detergent.
After you clean the Tester, do not use it for a period
of 5 hours.
When to Calibrate
Fluke recommends a calibration interval of 1 year.
Battery Replacement
If b (Fluke T90/T110) is on or B shows
in the LCD (Fluke T130/T150) during tests or
measurements, replace the batteries.
To replace the batteries:
1. Disconnect the Tester from the measurement
circuit.
2. Open the battery cover. See Figure 8.
3. Remove the discharged batteries.
4. Replace with two new 1.5V IEC LR03 AAA
batteries.
5. Align the battery polarity as shown on the case
housing.
6. Close and attach the battery cover.
Note
Do not overtighten the screw for the battery cover.
7. Complete a self‑test.
8
Содержание
- Instruction sheet 1
- T90 t110 t130 t150 1
- Voltage continuity tester 1
- Beeper t110 t130 t150 3
- Continuity diode test 3
- Display hold t130 t150 3
- Maintenance 3
- Resistance test t150 3
- Rotary field indication t110 t130 t150 3
- Single pole phase test 3
- Torch and backlight t110 t130 t150 3
- Voltage test 3
- Voltage test with switched load rcd trip test t110 t130 t150 3
- Battery replacement 4
- How to clean 4
- Limited warranty limitation of liability 4
- Specifications 4
- When to calibrate 4
Похожие устройства
- Gigabyte H310M S2V 2.0 (rev. 1.0) Инструкция
- Gigabyte H310M S2V (rev. 1.0) Инструкция
- Gigabyte H310M S2P 2.0 (rev. 1.0) Инструкция
- Fluke FLK-T5-100-KIT Инструкция по эксплуатации
- Fluke FLK-T5-100-KIT Инструкция EN
- Fluke FLK-T5-100-KIT Документация
- Gigabyte H310M S2P (rev. 1.0) Инструкция
- Fluke FLK-T5-100-KIT Расширенная инструкция
- Gigabyte H310M S2H 2.0 (rev. 1.0) Инструкция
- Fluke FLK-T5-1000 Инструкция по эксплуатации
- Fluke FLK-T5-1000 Документация
- Fluke FLK-T5-1000 Инструкция EN
- Fluke FLK-T5-1000 Расширенная инструкция
- Gigabyte H310M S2H (rev. 1.1) Инструкция
- Fluke FLK-T5-600 Инструкция по эксплуатации
- Fluke FLK-T5-600 Инструкция EN
- Fluke FLK-T5-600 Документация
- Fluke FLK-T5-600 Расширенная инструкция
- Fluke FLK-T6-1000 Инструкция по эксплуатации
- Fluke FLK-T6-1000 Документация