States at delivery
C.A 6522 and C.A 6524 delivered with a carrying bag for hands-free use, 2 elbowed-straight safety
leads (red and black) 1.50 m long, one red crocodile clip, one black test probe, 6 x LR6 or AA
batteries, multilingual user’s manual on CD, a quick start-up guide, multilingual safety datasheet.
C.A 6526, ditto + a CD containing
the Megohmmeter Transfer software.
C.A 6532 and C.A 6534, same as C.A 6526
+ 2wire-grips (red and black).
C.A 6536, same as C.A6524
+ 2wire-grips (red and black).
References
C.A 6522 megohmmeter .............................................................. P01140822
C.A 6524 megohmmeter .............................................................. P01140824
C.A 6526 megohmmeter .............................................................. P01140826
C.A 6532 megohmmeter .............................................................. P01140832
C.A 6534 megohmmeter .............................................................. P01140834
C.A 6536 megohmmeter .............................................................. P01140836
Accessories
Type-3 remote control probe ........................................................ P01102092A
Continuity rod ............................................................................. P01102084A
Thermometer + K thermocouple, C.A 861 ...................................... P01650101Z
C.A 846 thermo-hygrometer ......................................................... P01156301Z
USB-Bluetooth adapter ................................................................. P01102112
DataView
®
software...................................................................... P01102095
Replacement parts
2 elbowed-straight safety leads (red and black) 1.50 m long ........... P01295453Z
2 crocodile clips (red and black) .................................................... P01295457Z
2 test probes (red and black) ........................................................ P01295454Z
2 wire-grips (red and black) .......................................................... P01102053Z
Carrying bag for hands-free use ................................................... P01298049
C.A 6522 C.A 6524 C.A 6526 C.A 6532 C.A 6534 C.A 6536
Voltage
Measurement range / Resolution 0.3 V - 399.9 V / 0,1 V; 400 V - 700 V / 1 V
Accuracy / Input impedance ± (3 % + 2 cts) / 400 kΩ
Operating frequency DC; 15.3 - 800 Hz
Frequency
Measurement range / Resolution / Accuracy 15.3 Hz - 399.9 Hz / 0.1 Hz / ± (1 % + 2 cts); 400 - 800 Hz / 1 Hz / ± (1 % + 1 ct)
Insulation
Test voltage 250 - 500 - 1,000 V 50 - 100 - 250 - 500 - 1,000 V 50 - 100 V 10 - 25 - 100 - 250 - 500 V 10 V to 100 V in 1 V steps
Range at maximum test voltage 40 GΩ 200 GΩ 20 GΩ 50 GΩ 20 GΩ
Compliance with IEC 61557-2 standard 2 GΩ 2 GΩ 2 GΩ
Measurement range: 10 V 2 kΩ - 1 GΩ 2 kΩ - 2 GΩ
25 V 5 kΩ - 2 GΩ
(U
N
/5) kΩ -
(U
N
/5) GΩ
50 V 10 kΩ - 10 GΩ 10 kΩ - 10 GΩ
100 V 20 kΩ - 20 GΩ 20 kΩ - 20 GΩ 20 kΩ - 10 GΩ 20 kΩ - 20 GΩ
250 V 50 kΩ - 10 GΩ 50 kΩ - 50 GΩ 50 kΩ - 25 GΩ
500 V 100 kΩ - 20 GΩ 100 kΩ - 100 GΩ 100 kΩ - 50 GΩ
1,000 V 200 kΩ - 40 GΩ 200 kΩ - 200 GΩ
Variable test voltage 10 V to 100 V
Measurement range / Resolution
10
(1)
- 999 kΩ and 1.000 - 3.999 MΩ / 1 kΩ; 4.00 - 39.99 MΩ / 10 kΩ; 40.0 - 399.9 MΩ / 100 kΩ;
400 - 3,999 MΩ / 1 MΩ; 4.00 - 39.99 GΩ / 10 MΩ; 40.0 - 200 GΩ / 100 MΩ
Accuracy ± (3 % + 2 cts)
(2)
± (3 % + 2 cts)
(3)
Test voltage (I < 1 mA) - 0 % + 20 % ± 0.5 V
Display of test voltage ± (3 % + 3 cts) ± (3 % + 3 cts)
Display of test current / resolution 0.01 µA - 39.99 µA / 10 nA; 40.0 - 399.9 µA / 100 nA; 0.400 - 2.000 mA/ 1 µA
Accuracy of test current ± (10 % + 3 cts)
PI/DAR ratios 10 min / 1 min - 1 min / 30 s
Timer (min:s) 0:00 - 39:59
Discharge time (at 25V) < 2 s/µF
Alarms 2 fixed thresholds + 1 programmable threshold
Continuity
Range for continuity measurement 0.00 Ω - 10.00 Ω (200 mA) 0.00 Ω - 10.00 Ω (200 mA); 0.0 - 100.0 Ω (20 mA)
Accuracy / Open-circuit voltage ± (2 % + 2 cts) / ≥ 6 V
Measurement current 200 mA range: 200 mA (-0 mA +20 mA); 20 mA range: 20 mA ± 5 mA
Continuity thresholds (quick beep) 2 Ω fixed 2 Ω, 1 Ω, programmable threshold
Compensation of cables up to 9.99 Ω
Resistance
Measurement range / Resolution / Accuracy
0 - 3,999 Ω / 1 Ω; 4.00 kΩ - 39.99 kΩ / 10 Ω / ± (3 % + 2 cts)
40.0 kΩ - 399.9 kΩ / 100 Ω; 400 kΩ - 1,000 kΩ / 1 kΩ / ± (3 % + 2 cts)
Capacitance
Measurement range / Resolution
0.1 nF - 399.9 nF / 0.1 nF
400 nF - 3,999 nF / 1 nF
4.00 µF - 10.0 µF / 10 nF
Accuracy ± (3 % + 2 cts)
Line length 0 - 100 km
General specifications
Display 2 x 4,000 cts + logarithmic bargraph
Storage 300 measurements 1,300 measurements
Communication Bluetooth
®
Class II
Power supply / Automatic shutdown 6 x LR6(AA) batteries / 5 min, deactivatable
Battery life
1,500/2,500/6,000
measurements
(4)
: U
N
x 1 kΩ @ U
N
(5 s ON / 55 s OFF); 3,000 continuity measurements (5 s ON / 55 s OFF)
Dimensions (h x w x d) / Weight / IP rating 211 x 108 x 60 mm / 850 g / IP 54 / IK 04
EMC / Electrical safety IEC 61326 - 1 / IEC 61010 - 1 and IEC 61010 - 2 - 030, 600 V CAT IV
Compliance with standards IEC 61557 parts 1, 2, 4 and 10
Insulation and Continuity Testers
(1): 2 kΩ for C.A 6532 - C.A 6534 - C.A 6536 – (2): to be added: 10 V: 1 % per 0.1 GΩ ; 25 V: 0.4 % par 0.1 GΩ , 50 V: 2 % per GΩ, 100 V: 1 % per GΩ ; 250 V: 0.4 % per GΩ; 500 V: 0.2 % per GΩ; 1,000 V: 0.1 % per GΩ
(3): 10 % /U
N
is added per 100 MΩ – (4): depending on model.
Example for C.A 6532
FRANCE
Chauvin Arnoux
190, rue Championnet
75876 PARIS Cedex 18
Tel: +33 1 44 85 44 38
Fax: +33 1 46 27 95 59
export@chauvin-arnoux.fr
www.chauvin-arnoux.com
UNITED KINGDOM
Chauvin Arnoux Ltd
Unit 1 Nelson Ct, Flagship Sq, Shaw Cross Business Pk
Dewsbury, West Yorkshire - WF12 7TH
Tel: +44 1924 460 494
Fax: +44 1924 455 328
info@chauvin-arnoux.co.uk
www.chauvin-arnoux.com
MIDDLE EAST
Chauvin Arnoux Middle East
P.O. BOX 60-154
1241 2020 JAL EL DIB - LEBANON
Tel: +961 1 890 425
Fax: +961 1 890 424
camie@chauvin-arnoux.com
www.chauvin-arnoux.com
Содержание
- C a 6522 c a 6524 c a 6526 c a 6532 c a 6534 c a 6536 1
- Insulation and continuity testers 1
- Accessories software 2
- Activation of remote tests real time display dar and pi references graphical plot of the tests generation of measurement reports 2
- Chauvin arnoux s compact lightweight testers are easy to learn how to use all the information and measurement results are displayed on the backlit screen 2
- Equipped with a torch function this probe effectively lights the point of measurement 2
- Ergonomics 2
- Insulation and continuity testers 2
- Insulation measurement can be triggered with a single press on the probe s yellow button 2
- Software with its megohmmeter transfer module automatically recognizes the instrument when it is connected to the pc and opens the corresponding menu this tree structure menu offers users direct access to the data recorded in the instrument and its configuration the other functions available are 2
- Thanks to the tester s backlit screen the measurement results are visible at all times including in places with poor lighting 2
- The remote control probe is an accessory specially developed to make your insulation tests in the field simpler and quicker 2
- Applications 3
- C a 6522 c a 6524 and c a 6526 3
- C a 6532 3
- C a 6536 3
- Electronics c a 6534 3
- Function c a 6522 c a 6524 c a 6526 c a 6532 c a 6534 c a 6536 3
- Ideal for measurements on telephone lines insulation tests at 50 v or 100 v specific functions measurement of resistance capacitance leakage current and ac voltage measurement of the difference in resistance between 2 wires in a pair by using the δrel function display of the length of the line being tested thanks to programming of the linear capacitance in nf km 3
- Industrial maintenance 3
- Measurement of insulation over a programmed duration alarms and pass fail indicator light c a 6526 pi and dar ratios to determine the quality of the insulation with the advantage of not being significantly influenced by the temperature storage in memory to compare the measurement histories 3
- Periodic insulation tests on installations and equipment help to prevent incidents by setting up preventive maintenance designed to detect ageing and premature deterioration of their insulation characteristics 3
- Special 3
- Telecommunications 3
- Thanks to its wide test voltage range from 10 v to 500 v this model covers the specific requirements of the electronics industries for both low current applications and power stages its measurement range is from 2 kω to 50 gω with appropriate electrodes the 10 v and 100 v test voltages can also be used for electrostatic discharge applications 3
- With a variable test voltage from 10 v to 100 v in 1 v increments this model is intended for special applications in sectors such as avionics space and defence which often require a low test voltage and strict adherence to the programmed value 3
- Accessories 4
- C a 6522 c a 6524 c a 6526 c a 6532 c a 6534 c a 6536 4
- Insulation and continuity testers 4
- References 4
- Replacement parts 4
- States at delivery 4
Похожие устройства
- Komfovent Domekt CF 300 V Паспорт и инструкция по монтажу
- CHAUVIN ARNOUX CA-6528 Инструкция язык en
- CHAUVIN ARNOUX CA-6528 Data Sheet
- Komfovent Domekt CF 300 V Руководство пользователя
- Komfovent Domekt CF 300 V Технические данные
- Komfovent Domekt CF 400 V Технические данные
- Komfovent Domekt CF 400 V Паспорт и инструкция по монтажу
- Komfovent Domekt CF 400 V Руководство пользователя
- Komfovent Domekt CF 700 V Технические данные
- Komfovent Domekt CF 700 V Паспорт и инструкция по монтажу
- Komfovent Domekt CF 700 V Руководство пользователя
- Daikin EKAC200A Сервис мануал
- Daikin EKAC200A Инструкция по монтажу
- Komfovent Domekt CF 700 H Технические данные
- Komfovent Domekt CF 700 H Паспорт и инструкция по монтажу
- Komfovent Domekt CF 700 H Руководство пользователя
- EXTECH EX380260 Инструкция по эксплуатации
- EXTECH EX380260 Data Sheet
- EXTECH EX380260 Расширенная инструкция
- EXTECH EX380260 Инструкция язык en