MEAN WELL RPS-45-7.5 [2/4] W single output medical type

MEAN WELL RPS-45-3.3 [2/4] W single output medical type
VOLTAGE TOLERANCE
Note.4
I/P-O/P: 4KVAC
VOLTAGE ADJ.RANGE
3.8~5V 5.7~6.8V 8.62~11.3V 13.8~16.2V 17.25~20.3V 28.4~32.4V 55.2~64.8V
SPECIFICATION
ORDER NO.
DC VOLTAGE
RATED CURRENT
CURRENT RANGE
RATED POWER
RIPPLE & NOISE (max.)
Note.3
OUTPUT
VOLTAGE RANGE
Note.5
FREQUENCY RANGE
EFFICIENCY (Typ.)
AC CURRENT (Typ.)
INPUT
ENVIRONMENT
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT(max.)
SAFETY STANDARDS
SAFETY &
EMC
(Note. 7)
WORKING TEMP.
WORKING HUMIDITY
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
MTBF
DIMENSION
OTHERS
NOTE
PACKING
PROTECTION
RPS 45 3 3--.
3.3V
80mVp-p
LINE REGULATION
LOAD REGULATION
±0.5%
±2.0%
80 ~ 264VAC
47 ~ 63Hz
80%
1.2A / 115VAC 1A / 230VAC
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
I/P-O/P:100M Ohms / 500VDC / 25 / 70% RH
726.2Khrs min.
-30 ~ +70 (Refer to "Derating Curve")
20% ~ 90% RH non-condensing
-40 ~ +85 , 10 ~ 95% RH
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
76.2*50.8*24mm (L*W*H)
0.11Kg; 120pcs/14.2Kg/0.97CUFT
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2 medical level, criteria A
OVERLOAD
OVER VOLTAGE
115 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
8A
0 ~ 8 8A
.
26.4W
Protection type : Shut down o/p voltage, re-power on to recover
ANSI/AAMI ES60601-1 / TUV EN60601-1 / EС60601-1 approvedI
Compliance to EN55011(CISPR11) class B, EN61000-3-2,3
SETUP, RISE TIME
HOLD UP TIME (Typ.)
500ms, 30ms / 230VAC 500ms, 30ms / 115VAC at full load
30ms / 230VAC 12ms / 115VAC at full load
EMC IMMUNITY
EMC EMISSION
±0.03% / (0 ~ 50 )
℃℃
ISOLATION LEVEL
Primary-Secondary: 2xMOPP
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
3. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
4. Tolerance : includes set up tolerance, line regulation and load regulation.
5. Derating may be needed under low input voltages. Please check the derating curve for more details.
6. Touch current was measured from primary input to DC output.
The ambient temperature derating of 5 /1000m is needed for operating altitude greater than 2000m (6500ft).7.
8. The power supply is considered a component which will be installed into a final equipment. "All the EMC tests are been executed by mounting
the unit on a 360mm*360mm metal plate with 1mm of thickness." The final equipment must be re-confirmed that it still meets EMC directives.
For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
PEAK LOAD(10sec.) Note.2
29W
3.1~3.6V
±2.0%
Touch current< 100 A/264VAC
RPS 45--5
5V
80mVp-p
±0.5%
±2.0%
83%
8A
0 ~ 8 8A
.
40W
44W
47 55.~.V
±2.0%
RPS 45--.75
7.5V
80mVp-p
±0.5%
±2.0%
85%
0 ~ 5.95A
40.5W
44.6W
7.12~8.3V
±2.0%
RPS 45--12
12V
120mVp-p
±0.5%
±2.0%
88%
5.4A
0 ~ 4.18A
11.4~13.2V
±2.0%
RPS 45--15
15V
150mVp-p
±0.5%
±1.0%
89%
3.8A
0 ~ 3 3A
.
13.5~16.5V
±1.0%
RPS 45--24
24V
240mVp-p
±0.5%
±1.0%
90%
1.9A
0 ~ 2.1A
22.8~27.6V
±1.0%
RPS 45--48
48V
300mVp-p
±0.5%
±1.0%
91%
0.94A
0 ~ 1.03A
45.6~52.8V
±1.0%
3A
45.6W
50.2W
45W
49.5W
45.6W
50.2W
45.1W
49.4W
File Name: -SPEC 2015-11-04RPS 45-
45W Single Output Medical Type
RPS-45 series
СOLD STAR 30A/115VAC 60A/230VAC
MIL-HDBK-217(25 )