MEAN WELL RPS-500-24-SF [3/10] Rps 50
![MEAN WELL RPS-500-24 [3/10] Rps 50](/views2/1673348/page3/bg3.png)
RPS-500 series
SPECIFICATION
ISOLATION LEVEL
SAFETY STANDARDS
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
IEC60601-1, TUV EN60601-1,
I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25 / 70% RH℃
SAFETY &
EMC
(Note 8)
EMC IMMUNITY
EMC EMISSION
EN55024 , EN60601-1-2, EN61204-3
Parameter
Parameter
ESD
Standard
EN55011 (CISPR11)
EN55011 (CISPR11)
EN61000-3-2
EN61000-3-3
Test Level / Note
Class A
Class A
-----
100% dip 1 periods, 30% dip 25 periods,
Standard
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Test Level / Note
100% interruptions 250 periods
MTBF
DIMENSION
OTHERS
PACKING
194.1Khrs min. MIL-HDBK-217F (25 )℃
L*W*H
G.W.
Q’TY
P.W.
RPS-500
127x76.2x40mm
0.46Kg
30pcs
14.8Kg
RPS-500-C
130x86x43mm
0.54Kg
24pcs
14Kg
RPS-500-TF
130x86x66.5mm
0.58Kg
24pcs
14.9Kg
RPS-500-SF
160x86x43mm
0.64Kg
24pcs
16.4Kg
NOTE
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. Derating may be needed under low input voltages. Please check the derating curve for more details.
5. Touch current was measured from primary input to DC output.
6. The ambient temperature derating of 3.5 /1000m with fanless models and of 5 /1000m with fan models for operating altitude higher ℃℃
than 2000m(6500ft).
7. Please refer to Derating curve.
8. T EMC tests are he power supply is considered a component which will be installed into a final equipment All.
* . - executed by mounting the unit on a 360mm 360mm metal plate with 1mm of thickness The final equipment must be re confirmed that it still
.meets EMC directives For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com).
UL ANSI / AAMI ES60601-1 (3.1 version),EAC TP TC 020.
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved;
Type
M’MENT
1.03CUFT
0.77CUFT
0.86CUFT 0.91CUFT
Design refer to EN60335 1 (By request)-
5x3x157inch″″.″
5 11 x3 39 x1.69 inch.″.″ ″
511x339x262inch.″.″.″
6 3 x3 39 x1.69 inch.″ . ″ ″
μ μ
RF field susceptibility
EFT bursts
Surge susceptibility
Conducted susceptibility
Magnetic field immunity
Voltage dip, interruption
Conducted emission
Radiated emission
Harmonic current
Voltage flicker
Level 4, 30A/m
Level 3, 10V
Level 3, 2KV
Level 4, 15KV air ; Level 4, 8KV contact
Level 4, 4KV/Line-FG ; 2KV/Line-Line
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
File Name:RPS-500-SPEC 2019-06-06
500W Reliable Green Medical Power Supply5"×3"
Class : Class B ClassⅠⅡ, Class : A
EMI Performance
Conducted Radiated
ClassⅠ
Class B
(with FG)
Class Ⅱ
(no FG)
Class A
Class A
Class A
Содержание
- Rps 50 1
- Series 1
- W reliable green medical power supply 1
- Rps 50 2
- Series 2
- W reliable green medical power supply 2
- Design refer to en60335 1 by request 3
- Rps 50 3
- Series 3
- W reliable green medical power supply 3
- Block diagram 4
- Output derating vs input voltage 4
- Rps 50 4
- Series 4
- W reliable green medical power supply 4
- Ambient temperature 5
- Convection 5
- Force air 5
- Order no 5
- Products 5
- Rps 50 5
- Rps 500 5
- Rps 500 c 5
- Rps 500 sf 5
- Rps 500 tf 5
- Series 5
- W reliable green medical power supply 5
- With 25cfm fan 5
- Air flow direction 6
- Bottom view 6
- Rps 50 6
- Series 6
- W reliable green medical power supply 6
- Rps 50 7
- Series 7
- W reliable green medical power supply 7
- Air flow direction 8
- Bottom view 8
- M3 l 2 8
- M3 l 4 8
- Rps 50 8
- Series 8
- Side view 8
- Top view 8
- Unit mm case no 247a d 247b t 8
- W reliable green medical power supply 8
- Air flow direction 9
- Rps 50 9
- Series 9
- W reliable green medical power supply 9
- Ac input connector cn1 jst b3p vh or equivalent 10
- Dc output connector cn2 cn3 10
- Fan connector cn12 tkp 8812 2 or equivalent except for rps 500 tf sf 10
- Function connector cn11 tkp dh2i 2x2 or equivalent 10
- Function connector cn95 tkp dh2l 2x2 or equivalent 10
- Rps 50 10
- Series 10
- W reliable green medical power supply 10
Похожие устройства
- MEAN WELL RPS-500-24-SF Каталог
- MEAN WELL RPS-500-24-TF Инструкция по эксплуатации
- MEAN WELL RPS-500-24-TF Data Sheet
- MEAN WELL RPS-500-24-TF Каталог
- MEAN WELL RPS-500-24-TF Инструкция язык en
- MEAN WELL RPS-500-27 Инструкция по эксплуатации
- MEAN WELL RPS-500-27 Инструкция язык en
- MEAN WELL RPS-500-27 Data Sheet
- MEAN WELL RPS-500-27 Каталог
- MEAN WELL RPS-500-27-C Инструкция по эксплуатации
- MEAN WELL RPS-500-27-C Инструкция язык en
- MEAN WELL RPS-500-27-C Data Sheet
- MEAN WELL RPS-500-27-C Каталог
- MEAN WELL RPS-500-27-SF Инструкция по эксплуатации
- MEAN WELL RPS-500-27-SF Инструкция язык en
- MEAN WELL RPS-500-27-SF Data Sheet
- MEAN WELL RPS-500-27-SF Каталог
- MEAN WELL RPS-500-27-TF Инструкция по эксплуатации
- MEAN WELL RPS-500-27-TF Инструкция язык en
- MEAN WELL RPS-500-27-TF Data Sheet