Tanita KP-601 [10/13] Output data format usb verison only

Tanita KP-601 [10/13] Output data format usb verison only
8
TECHNICAL SPECIFICATIONS
Test weight 100 g (included)
Capacity 100.000 g / 500.00 ct / 3.5273 oz /3.2150 ozt / 64.300 dwt /
1543.25 grain / 2.6717 HT / 2.6666 TT
Graduation 0.002 g / 0.01 ct / 0.0001 oz / 0.0001 ozt / 0.002 dwt / 0.05 grain /
0.0001 HT / 0.0001 TT
Power DC 6V AA (LR06) x 4 (included)
AC adaptor (DC 6V 200 mA) (Option)
Power Consumption
(MAX)
Dual LCD: 45 mA, single LCD: 40 mA
*Not legal for trade
OUTPUT DATA FORMAT (USB verison only)
This section covers the exporting of data from the scale to an external device (e.g. PC) using a
USB compliant signal.
• USB interface is for data OUTPUT ONLY!
• This scale is not capable of receiving instructions from an external device.
• This scale will only output data from 0.04g to 100.00g.
Specifi cations
Communications standard USB
Communications method Asynchronous transaction
Signal speed 9600 baud
Data bit length 8 bits
Parity None
Stop bit 1 bit
Terminator CR+LF
Note:
USB connecter (mini-B-type 4-pin female) is located on the side of the scale.
Caution
If you want to use the USB output, you must install the necessary driver onto your
PC, available to download from http://www.tanita.com
USB Port