Shimadzu UW-4200H [57/129] Calibration check using the built in weight uw series only

Shimadzu UW-2200H [57/129] Calibration check using the built in weight uw series only
41
10. Calibration
Read Chapter 7 Before Use of These Pages
10.2.2
Calibration Check Using the Built-in Weight (UW Series Only)
The shift from the last calibration is displayed using the built-in calibration weight, however, the balance
is not adjusted.
1 Verify that the balance is in mass display and
that the pan is empty.
2 Press the [CAL] key once to display “i-tESt”.
(If “i-tESt” is not displayed, return to mass display
and select menu item .)
3 Press the [O/T] key.
The display changes sequentially from “i-tESt 2” to
the “d xxx” display. (xxx indicates a numeric value)
This “d” value indicates the difference between
the current calibration weight reading and the cal-
ibration weight reading at the last span calibration.
4 To perform span calibration, change the “d”
value to zero, by pressing the [CAL] key.
--Otherwise, Press the [O/T] key to avoid chang-
ing the “d” value to zero. (Pressing the [POWER]
key interrupts calibration and does not change
this value to zero.)
“CALEnd” is displayed, indicating the completion
of the calibration check.
2
Note
Changing the “d” value to zero is equivalent to performing span calibration.
Notes
Examples for interpreting the results of a Calibration Check:
Error codes that may be displayed:
“d ouEr” (d OVER) indicates that the “d” value is 1000 counts or more.
“d UndEr” indicates that the “d” value is -1000 counts or less.
“d” Value Actual Mass Displayed Mass
0.3 3000g 2999.7(3200g/0.1g balance)
+0.21 400g 400.21(420g/0.01g balance)