Elcold PRO 21 [16/18] Factory settings pro xlt models

Elcold PRO 21 [16/18] Factory settings pro xlt models
User Instructions Low Temperature Freezers Model PRO/XLT
Factory settings PRO/XLT models.
Technical specifications:
15
Ambient temperature
Net volume
Net volume
Height with 50 mm castors
Height with open lid
Width
Insulation thickness
Depth excl. handle and hinges
Power consumption at +25˚ C
Fuse
Temeprature range
Noise level
Weight
˚C
litres
cu. ft.
mm
mm
mm
mm
mm
Watts
A
˚C
dBa
kg
+15˚ to +25˚
130
4,6
920
1620
720
100
730
330
10
-30˚ to -60˚
< 51
53
+15˚ to +25˚
230
8,1
920
1620
1050
100
730
410
10
-30˚ to -60˚
< 51
68
+15˚ to +25˚
300
10,6
920
1620
1300
100
730
560
10
-30˚ to -60˚
< 51
78
PRO 11/11 XLT PRO 21/21 XLT PRO 31/31 XLT
PRO/XLT
PRO 60/60 XLTPRO 41/41 XLT
+15˚ to +25˚
360
12,8
920
1620
1500
100
730
610
10
-30˚ to -60˚
< 51
89
+15˚ to +25˚
485
17,3
920
1620
1700
80
730
610
10
-30˚ to -60˚
< 51
95
Description
Regulation
Set point
Different (Hysteresis)
Minimum set point
Maximum set point
Action type output 1
Anti-short cycle delay
Minimum time a stage stays switched on
Minimum time between 2 following
switching ON of the same load
Alarm and safety
Alarm configuration
Minimum alarm
Maximum alarm
Alarm recovery differential
Alarm delay
Alarm delay at start up
Alarm relay disabling
Alarm relay polarity
Out1 status with faulty pr.
Other
Probe calibration
Resolution
Measurement unit (temp.)
Kind of probe
3rd. wire presence
Energy saving differencial
Digital input configuration
Digital input polarity
Alarm delay for dig. input
Serial address
oFF function enabling
Label
Set
Hy1
LS1
US1
S1C
Ac
on
ono
-60
Food
-60
1,5
-60
-45
in
0
0
0
ALC
ALL
ALU
ALH
Ald
dAO
tbA
AS
So1
Opb
rES
udM
Pbc
P3F
HES
i1F
i1P
did
Adr
OnF
re
5,0
20,0
2,0
60
10,0
on
OP
OFF
5
De
C
Pt
YES
0
EAL
cL
0
1
NO
Dixell XT111C
-65
Research
-65
0,7
-65
-45
in
0
0
0
re
3,0
10,0
2,0
5
8,0
on
OP
OFF
2
De
C
Pt
YES
0
EAL
cL
0
1
NO
Скачать