Supplement 712 Instruction Sheet
2 4/09
3. To disable, press until the display shows
OFF.
Span Check
The calibrator allows you to store 0% and 100%
setpoints for each output type. Once setpoints are
stored, the span check feature allows you to quickly
toggle back and forth from 0% to 100% or to step in
25% increments.
Automatic step and ramp modes can be enabled
while in span check mode by simultaneously
pressing or . First select the desired output
range, then proceed to store the setpoints:
1. Use and to set the output to the desired
value for 0%.
2. Press and simultaneously to store the
0% value.
3. Use and to set the output to the desired
value for 100%.
4. Press and simultaneously to store the
100% value.
Under Testing and Replacing the Fuses, delete the
entire section and the corresponding figure.
Under Replacement Parts and Accessories, delete
the F1, F2 row and under MP86 change the part
number,
From: 620168
To: 2397526
Remove the F1 and F2 fuses from the replacement
parts illustration.
Under Specifications, replace the Ohms
Specifications table with the following two tables: