Safety Information
2
XW Warning
To avoid electric shock, injury, or damage to the
Calibrator:
Use the Calibrator only as described in this
User Manual and the Fluke 718Ex CCD
(Concept Control Drawing) or the protection
provided by the Calibrator may be impaired.
Inspect the Calibrator before use. Do not use
it if it appears damaged.
Check the test leads for continuity, damaged
insulation, or exposed metal. Replace
damaged test leads.
When using probes, keep fingers behind the
finger guards on the probes
Never apply more than 30.0 V between the
input terminals, or between any terminal and
earth ground.
Applying more than 30.0 V to the input
terminals invalidates the Calibrator’s Ex
Approval and may result in permanent
damage to the unit so it can no longer be
used.
Use the proper terminals, mode, and range for
the measuring or sourcing application.
To prevent damage to the unit under test, be
sure the Calibrator is in the correct mode
before connecting the test leads.
When making connections, connect the COM
test probe before the live test probe. When
disconnecting, disconnect the live probe
before the COM probe.
Never use the Calibrator with the red holster
removed.
Never open the Calibrator case. Opening the
case invalidates the Calibrator’s Ex Approval.
Make sure the battery door is closed before
using the Calibrator.
Replace the battery as soon as B (low
battery) appears to avoid false readings that
can lead to electric shock. Remove the
Calibrator from the Ex-hazardous area before
opening the battery door.
Remove test leads from the Calibrator before
opening the battery door.
This equipment is specified for use in
measurement category I (CAT I) pollution
degree 2 environments and should not be
used in CAT II, CAT III, or CAT IV
environments. Voltage transients should not
exceed 300 volts for the CAT I applications
where this product is used. Measurement
transients are defined in IEC1010-1 as 2 µs
rise time with a 50 µs duration at 50 % of the
maximum amplitude height.
Measurement Category I (CAT I) is defined for
measurements performed on circuits not
directly connected to the mains.
Turn off circuit power before connecting the
Calibrator mA and COM terminals in the
circuit. Place Calibrator in series with the
circuit.
When servicing the Calibrator, use only
specified replacement parts. Do not open the
Calibrator case. Opening the case invalidates
the Calibrator’s Ex Approval.
Do not allow water inside the case.
Do not use in a damp or wet environment.
When using the Calibrator’s internal pressure
sensor, do not connect a pressure module at
the Calibrator to avoid misleading readings. If
both a pressure module and the internal
pressure sensor are connected, the Calibrator
displays ONLY the pressure module
measurement. To avoid misleading readings,
disconnect the pressure module connector at
the Calibrator.