Pressure Calibrator
Safety Information
3
To avoid a violent release of pressure in a
pressurized system, shut off the valve and
slowly bleed off the pressure before attaching
or detaching the internal pressure sensor or
pressure module fitting to the pressure line.
To avoid overpressure damage, do not apply
pressure to the internal pressure sensor input
that exceeds the following:
Model 718Ex 30G: 30.000 psi, 206.85 kPa,
or 2.0685 bar. OL appears at 33 psi.
Model 718Ex 100G: 100.00 psi, 689.5 kPa,
or 6.895 bar. OL appears at 120 psi.
Model 718EX 300G: 300.00 psi, 2068 kPa,
or 20.68 bar. OL appears at 360 psi.
When measuring the pressure of potentially
hazardous gases, care must be taken to
minimize the possibility of leakage:
Confirm that all pressure connections are
properly sealed.
Confirm that the Pressure/Vacuum
Release Control is in the closed position
(fully clockwise) and the
Pressure/Vacuum switch is in the +
position (fully clockwise).
If the Calibrator has been dropped or
subjected to rough handling, remove the
Calibrator to a safe area and check for
leaks to confirm the integrity of the
internal pneumatic components.
Do not use a Model 718Ex (including 718Ex
300G) to measure potentially hazardous gases
at pressure greater than 100 psi (6.9 bar).
Do not use in a damp or wet environment.
Faults and Damage
Applying a voltage greater than 30 V to the input of the Calibrator
invalidates its Ex Approval and may impair its safe operation in
an Ex-hazardous area.
If there is any reason to suspect that the safe operation of the
Calibrator has been affected, it must be immediately withdrawn
from use, and precautionary measures must be taken to prevent
any further use of the Calibrator in an Ex-hazardous area.
Fully observe all instructions, warnings, and cautions contained
in this manual. In case of doubt due to translation and/or printing
errors, refer to the original English users manual.
The safety features and integrity of the unit may be compromised
by any of the following:
External damage to the housing
Internal damage to the Calibrator
Exposure to excessive loads
Incorrect storage of the unit
Damage sustained in transit
Correct certification is illegible
Using the product with the red holster removed
Functioning errors occur
Permitted limitations are exceeded
Functioning errors or obvious measurement inaccuracies
occur which prevent further measurement by the Calibrator
Opening the case
Safety Regulations
The use of the Calibrator meets the requirements of the
regulations providing that the user observes and applies the
requirements as stated in the regulations and that improper and
incorrect use of the unit is avoided.
Use must be restricted to the specified application
parameters.
Do not open the Calibrator.
Do not remove or install the battery within the Ex-hazardous
area.
Do not carry additional batteries within the Ex-hazardous
area.
Use only type-tested batteries. The use of any other
batteries will invalidate the Ex-certification and present a
safety risk.