Fluke 718 ex [4/6] Change 5 60644

Fluke 718 ex [4/6] Change 5 60644
718Ex 30G/100G/300G Users Manual Supplement
10/17 3
Change #5, 60644
On page 25, replace the Pressure Sensor Input Table and add the Note:
Pressure Sensor Input
Model Range Accuracy
Max Non-destructive
Pressure
30G
-12 to 30 psi
(-83 to 207 kPa)
±0.025 % of range (6 month)
±0.035 % of range (1 year)
60 psi
(413 kPa)
100G
-12 to 100 psi
(-83 to 690 kPa)
200 psi
(1.4 MPa)
300G
-12 to 300 psi
(-83 to 2068 kPa)
±0.05 % of range
(1 year)
375 psi
(2.6 MPa)
Temperature coefficient: 0.01 % of range per °C for temperature ranges -10 °C to 18 °C and 28 °C to 55 °C
Note
Specifications apply to the 718Ex Pressure Calibrators with firmware version 2.0 or greater.
To verify the firmware version, press and hold
and power on the Calibrator.