ZHM 125B
MEASUREMENT FOUNDATION
17
Maintenance
Always keep instrument dry.
Prevent dirt from getting in between the measuring electrodes
Changing the electrodes:
Unscrew the 5 screws at the back of the instruments.
Remove housing lid. Unscrew all 5 screws on the electronic plate. Replace electrodes.
Close instrument again.
Function check
Connect electrodes with T contacts to the protective cap.
Reference display for wood: 27%±2%
Reference display for building material: 1.25%±0.1%
Function error: Send instrument for servicing.