Replace the fuse as follows:
1. Remove the test leads and turn the calibrator off.
2. Remove the battery door.
3. Remove the three Phillips-head screws from the case
bottom and turn the case over.
4. Gently lift the top cover from the end nearest the input/output
terminals until it unsnaps from the bottom cover.
5. Replace the fuse with a 0.125 A 250 V fast fuse, Littelfuse
®
2AG.
6. Fit the top and bottom covers together, engaging the two
snaps. Make sure that the gasket is properly seated.
Reinstall the three screws.
7. Replace the battery door.