Schmidt Hammer 225
MEASUREMENT FOUNDATION
3. Updated the concrete strength presumed method for single piece and batch inspecting.
Calibrate Test hammer
In order to ensure the Test Hammer reliability of technical performance and rebound method measurement accuracy
should be regular or irregular status of the instrument calibration. Calibration of test hammer divided into general and
periodic.
General calibration
Hammer commonly used frequently within effective certication period, the user as the case may be self-calibrate. The
standard self-calibrate devices at home and abroad are generally Hammer steel anvil at present For the Test Hammer
nominal kinetic energy < 2.207N. m, using GZ16 steel anvil Hammer machine technical status of a general self-calibrate
is entirely feasible.
Usually when the Test Hammer use frequently or has doubt for the instrument inspecting data, can calibrate on the steel
anvil, looked at the calibrating of rebound values Rm whether in the 8 0 + 2 or not . With only the calibration of steel anvil
Hammer rebound values to determine whether it is in normal is not complete. Operation must obtain the following proce-
dure that test hammer’s state of technology compliance with standards and using requirements:
1. Check the maximum static friction force between pointer slider and shaft in the machine should be within the 0.5-0.8N,
if it does not meet the requirements to adjust.
2. Check the length of the effective work of strike tension spring is 61.5mm,
usually can adjust strike tension spring different holes location on spring seat to
meet the technical requirements.
3. Check the tension spring stretched standard length is 75mm. The technical parameter can be measured which instru-
ment should be work state, that is in the instrument to measure the length. Users inspecting the parameter is difcult to
measure under lack of detecting condition of Test Hammer.