Guidance and manufacturer’s declaration - electromagnetic immunity
The JFT90 is intended for use in the electromagnetic environment specified below. The customer or the
user of the JFT90 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast transient/
burst
IEC 61000-4-4
2 kV for power supply
lines
1 kV for input/output lines
Not applicable
Mains power quality
should be that of a typi-
cal commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV line(s) to line(s)
2 kV line(s) to earth
Not applicable
Mains power quality
should be that of a typi-
cal commercial or
hospital environment.