Problem Solution
Large variation in room tempera-
ture.
Store the device for at least 30 min-
utes in a room with room temperature
(10 4 0 ° C / 5 0 10 4 ° F ) .
Room temperature below 10 °C or
o v e r 4 0 ° C ( < 5 0 ° F, > 1 0 4 ° F ) .
Room temperature must be between 10 °C
and 40 °C (<50 °F, >104 °F).
Faultless functionality cannot be
guaranteed.
Remove the battery for approx. 1 min-
ute and then reinsert. In the event of a
repeated error message, contact your
retailer or Customer Services.
(1) Forehead temperature mode:
The temperature recorded is
h i g h e r t h a n 4 2 . 2 ° C ( 1 0 8 ° F ) .
(2) Object temperature mode: The
temperature recorded is higher
t h a n 8 0 ° C ( 1 7 6 ° F ) .
Operate the thermometer only between
the specied temperature ranges. In the
event of a repeated error message, con-
tact your retailer or Customer Services.
(1) Forehead temperature mode:
The temperature recorded is
l o w e r t h a n 3 4 ° C ( 9 3 . 2 ° F ) .
(2) Object temperature mode: The
temperature recorded is lower
t h a n - 2 2 ° C ( - 7 . 6 ° F ) .
Operate the thermometer only between
the specied temperature ranges. In the
event of a repeated error message, con-
tact your retailer or Customer Services.
Self-test unsuccessful. Replace the batteries.
The batteries are empty. Replace the batteries.