AXIOMET AX-T520 Расширенная инструкция онлайн [11/13] 527868

AXIOMET AX-T520 Расширенная инструкция онлайн [11/13] 527868
11
20mA 10µA
200mA 100µA ±1.5 % of rdg ± 2 digits
10A 10mA ±2.0 % of rdg ± 5 digits
AC CURRENT
Range Resolution Accuracy
200µA 0.1µA ±2.0 % of rdg ± 3 digits
2mA 1µA
±1.0 % of rdg ± 3 digits
20mA 10µA
200mA 100µA ±1.8 % of rdg ± 3 digits
10A 10mA ±3.0 % of rdg ± 5digits
Overload Protection: F 200mA/250V fuse for 200µA to 200mA ranges
Frequency Range: 40Hz to 400Hz (for AC)
Response: Average, Calibrated in rms of sine wave
4.4 RESISTANCE
Range Resolution Accuracy
200 0.1 ± 0.8 % of rdg ± 3 digits
2K 1 ± 0.8 % of rdg ± 2 digits
20K 10 ± 0.8 % of rdg ± 2 digits
200K 100 ± 0.8 % of rdg ± 2 digits
2M 1K ± 0.8 % of rdg ± 2 digits
20M 10K ± 1.0 % of rdg ± 2 digits
200M 100K ± 5.0 % of (rdg-10digits)
± 10 digits
Скачать