6.9.4 Polarization Index [PI] Insulation Testing
Read and understand all operations and safety information contained in the Insulation Resistance
measurement section above and the Safety section before continuing.
1. Connect the meter to the device under test as shown in the Insulation test section above.
2. Select the desired output test voltage using the rotary function switch.
3. Use the PI/DAR button to access the PI test mode.
4. The display will show the PI icon when the PI mode is successfully accessed.
5. The test times (10 minutes : 1 minute) will show on the lower right.
6. Press the TEST button to start the test.
7. During the test the auxiliary display digits show the test time in seconds, the voltage alert
flashes on the upper left, and the ‘test voltage alert’ LED illuminates.
8. The test will stop automatically after 10 minutes.
6.9.5 Comparator Utility
Read and understand all operations and safety information contained in the Insulation Resistance
and Continuity measurement sections and the Safety sections before continuing.
In Comparator mode the meter compares Insulation Resistance or Continuity measurements to a
preset resistance value and displays PASS (measured value greater than reference value) or
FAIL (measured value lower than reference value).
1. Connect the meter to the device under test as shown in the Insulation Resistance or
Continuity test sections above.
2. Press the COMP button momentarily and the COMP icon will appear on the lower left, next to
the comparison resistance value.
3. Press and hold the SET button until the SET icon is shown.
4. Use the
button to step to the COMP screen (see the SETUP mode section below for
more details on the SETUP mode).
5. Use the arrow buttons to step to the desired reference value.
6. Press
to confirm setting.
7. Press and hold the SET button to return to the test mode.
8. Press COMP momentarily and the COMP icon will appear.
9. Press the TEST button to start the test. If the measurement is greater than the reference then
the test passes and PASS will be displayed; if the measured value is lower than the reference
then the test fails and FAIL is displayed.
10. The Insulation Resistance comparator preset options are: 500k, 1M, 2M, 5M, 10M, 20M, 50M,
100M, 200m, and 500M.
11. The Continuity comparator preset options are: 1, 2, 5, 10, and 20.