9. Technical data
Technical data for the OMRON NE-C24 Compressor with the OMRON V.V.T.
(Virtual Valve Technology) Nebuliser kit:
Result of cascade impactor measurements for particle size**
* Measured by OMRON HEALTHCARE Co., Ltd.
**Independently measured by Prof. Dr. Hiroshi Takano, Department of Chemical
Engineering and Materials Science, Faculty of Science and Engineering,
Doshisha University, Kyoto, Japan, according to EN 13544-1:2007
Notes:
Performance may vary with drugs such as suspensions or high viscosity. See
drug supplier’s data sheet for further details.
See web site of OMRON HEALTHCARE EUROPE to update technical
information.
URL: www.omron-healthcare.com
Particle Size: **MMAD Approx. 3µm
MMAD = Mass Median Aerodynamic Diameter
Medication Tank
Capacity:
7ml maximum
Appropriate Medication
Quantities:
2ml minimum - 7ml maximum
Sound: *Approx. 46dB
Nebulisation Rate: *Approx. 0.3ml/min (by weight loss)
Aerosol Output: **0.47ml (2ml, 1%NaF)
Aerosol Output Rate: **0.06ml/min (2ml, 1%NaF)
Cumulative % particle mass of sodium fluoride under size
0
10
20
30
40
50
60
70
80
90
100
1000110.1
Particle Size Dp (μm)
Cumulative Undersize %
Individual
Mean
MMAD 3.0 μm
NE-C24_RU.book Page 26 Thursday, October 25, 2012 1:50 PM