If the data of each cycle differ too much from each
other, a fourth measurement is performed before
the result is displayed. In rare cases when the
blood pressure is so unstable that even after four
measurements the data vary too much, an error
message is displayed. Refer to Error Messages and
Malfunctions (Section 7) at the end of this manual.
5.6. Discontinuing a measurement
If it is necessary to interrupt a measurement for any reason, for example if the patient feels unwell,
press the O/I button. The device immediately lowers the cuff-pressure and enters into sleep mode.
6. Memory handling and user setting
6.1 Viewing Measurements Saved in Memory
The blood pressure monitor automatically stores each of the
last 30 measurement values. To view the stored measurements
for a user, press or in sleep mode. The last measurement
taken for the user is displayed.
The user, measurement sequence number, date, and time
are displayed on the top. Data obtained in Average Mode are
indicatedby«MAM»intheupperrightcorner.
Press or to scroll through the list items. The first item in
the list is the average of all stored data.
Press OK and the display shows table of all data.
The first entry is the average. Data obtained in Average Mode
areindicatedbyan«A»attheveryright.
Press or to scroll through the list.
13
15
Countdown
timer
Measurement
number
9 MAM
3/03 15:48
sys
124 mmHg
dia
73 mmhg
pul
77 /min.
Measurement
number
3/03 15:48
No. sys dia pul
avg. 149 110 85
1 145 109 83
2 119 71 74 A
3 124 73 79
4 130 80 77 A