Shimadzu UX-820S [114/129] Appendices

Shimadzu UX-620H [114/129] Appendices
Appendices
98
Appendices
A-1. Menu Map
•Pressing [CAL] key moves to the next menu in the same hierarchy. ( in menu map)
•Pressing [O/T] key moves to the menu of one hierarchy down. ( in menu map)
When no menu exists in the menu of one hierarchy down, it is fixed.
•Pressing [POWER] key returns to the menu of one hierarchy up. ( direction in menu map)
Refer to 7.3 for menu item selection.
Important Note on Menu Item Selection
Even the desired menu item is reached and displayed, it is not yet set unless Stability
mark ( ) is illuminated with it. Do not fail to press [O/T] key to put Stability mark
before returning to the mass display.
(The mass display)
CAL Execution menu
Executes calibration.
WindowsDirect Setting
mark blinking
Menu Group 1 ([CAL] key three times from mass-display)
Sets type of calibration to execute
blinking Sets I-CAL. (Adjusting sensitivity
with the built-in weight) # 1
blinking Sets I-TEST. (Testing sensitivity
with the built-in weight) 2
blinking Sets E-CAL. (Adjusting sensitivity
with the external weight) #
*
3
blinking Sets E-TEST. (Testing sensitivity
with the external weight)
*
4
blinking
Fully-automatic calibration PSC On/Off
Sets PSC ON 5
Sets PSC OFF # 6
blinking
Fully-automatic calibration
Clock-CAL settings
time 1 7
time 2 8
time 3 9
blinking Executes calibration of the built-in
calibration weight
*
10
*
Continued
: Menu item number
# : Default settings
No.
*Not applicable to a verified balance as a legal measuring instrument in the EU

Содержание