Shimadzu UX-820S [40/129] Set up windowsdirect

Shimadzu UX-820S [40/129] Set up windowsdirect
24
6. WindowsDirect Function
6. WindowsDirect Function
6.1 Introduction: Experience it!
The UW/UX series balance can transfer data directly to a computer running MS-Excel, mass input win-
dow of analytical instrument software or other applications on Windows
®
* OS, as if the displayed value
were typed from the keyboard to the cursor position. This function is called WindowsDirect. As this func-
tion directly accesses the Windows
®
OS, communication software-installation troubles are eliminated.
A cable and a few simple settings are all that is needed to enable data transmission from the balance.
For bi-directional communication between the balance and the computer, software is required. Win-
dowsDirect does not send commands from the computer.
Only numerical values can be transferred through WindowsDirect.
6.2 Set Up WindowsDirect
Simple settings are made for the balance and the computer. Connection is by RS-232C cable (15.2.1)
specified by Shimadzu.
If bi-directional communication software is used: WindowsDirect function should be turned off. Set up the opti-
mal communication parameters for the software according to “15.3 Communication setting”.
6.2.1 Setting Up the Balance
1 Press the [MENU] key twice from the gram-dis-
play. “Setwin” appears.
2 Press the [O/T] key. Verify the stability mark is
illuminated with “Setwin” display. All the communi-
cation settings for WindowsDirect have been made.
3 Go to “STAND-BY” by pressing the [POWER]
key several times and unplug the AC adapter
from the balance. Unpluging the balance once is
necessary after the above setting.
Notes
Individual communication parameters can be changed at any time using the communications set-
tings menu. If WindowsDirect communication settings have been previously made, the (stability
mark) may appear with the “SEtwin” display even after communication settings are changed and
become invalid for WindowsDirect. To restore WindowsDirect optimal settings, first go to the "SEt-
win" display and remove the stability mark by pressing the [O/T] key. This restores the default com-
munication settings. Then, reset “SEtwin” following the procedure described in 6.2.1.
•Refer to 15.3 for details of communication settings.