Shimadzu UX-6200H [124/129] A 7 below weigh hook dimensions

Shimadzu UW-2200H [124/129] A 7 below weigh hook dimensions
Appendices
108
Cornerload Performance
1 Allow the balance to warm up sufficiently by
turning ON the power and leaving it at the gram-
display at least two hours before starting the per-
formance checks.
2 Use a weight that is approximately 1/4 of the
balance capacity and place it sequentially at five
different positions on the pan in the order shown.
Record the readings in each of the five positions.
3 If the difference between the readings at the
center position and each of the off-center posi-
tions (eccentric error) is within five counts, the
balance cornerload performance is considered
normal.
* One count corresponds to one minimum dis-
play unit of the balance model.
A-7. Below-Weigh Hook Dimensions
1
2 3
4
5
0.7
4.5
Below-weigh Hook
Holeφ20
Balance bottom
43
3
4.6
(Unit: mm)