Ada tempro-pocket [7/30] Tem pro pocket

Ada tempro-pocket [7/30] Tem pro pocket
TemPro-pocket
7
TROUBLE SHOOTINGS
Problems Causes Solutions
No display Dead battery Check and replace the battery
Display
Low battery Replace the battery
Display “LO” Target temperature is lower than range Choose target of temperature within the measuring range
Display“HI” Target temperature is higher than range Choose target of temperature within the measuring range
Extremely low accuracy
Target surface is smaller than measuring spot
According to D:S ratio, make sure target surface is
larger than measuring spot.
Emissivity of target is far different from thermometer’s.
MAINTENANCE
1. Lens protection: abrupt temperature changes will cause condensing, do not wipe the lens if it happens, clean the lens with care after
natural evaporation. Blow off loose particles using clean compressed air. Gently brush remaining debris away with camel hair brush.
Carefully wipe the surface with a moist cotton cloth.
2. Keep clean, avoid dropping and shaking, use only mild soap and damp cloth to clean the tool. Never let any liquid get inside the tool
or immerse any part of the tool into liquid.