COSMO 100
19
Info CAUSE REMEDY
204
Calculation overflow Repeat procedure
252
Temperature too high Cool down instrument
253
Temperature too low Warm up instrument
255
Receiver signal too weak Use target plate
256
Received signal too strong Use target plate (grey side)
257
Wrong measurement Use target plate (brown side)
258
Wrong initialization Switch on – off the instrument
Hardware error
Switch on/off the device several times and check if the symbol
still appears. If so please call your dealer for assistance.